Suppr超能文献

超声剪切波弹性成像评估:动物离体研究。

Assessment of ultrasound shear wave elastography: An animal ex-vivo study.

机构信息

Department of Ultrasound, Peking University People's Hospital, Beijing, People's Republic of China.

Department of Ultrasound, Beijing Tsinghua Changgung Hospital, School of Clinical Medicine, Tsinghua University, Beijing, People's Republic of China.

出版信息

J Appl Clin Med Phys. 2023 Apr;24(4):e13924. doi: 10.1002/acm2.13924. Epub 2023 Feb 2.

Abstract

OBJECTIVES

To explore the influence of the surrounding environment of the target tissue, lesion size, and rectangular sampling box size on shear wave speed (SWS).

METHODS

The tendon SWS was acquired ex-vivo. Then the tendons were dissected and buried in the couplant (gel) and evaluated by two-dimensional shear wave elastography (2D-SWE). Finally, the tendons were placed in the isolated muscles to simulate the intramuscular lesions, and their elasticity was tested under two rectangular sampling box conditions. The isolated complete liver SWS was acquired. Similarly, the large and small pieces of livers were cut out, placed in the muscles, and assessed by SWE under two rectangular sampling box conditions. The SWS acquired under different conditions was compared. Variability was evaluated using the coefficient of variation (CV). The intraclass correlation coefficient (ICC) was used to evaluate repeatability.

RESULTS

The SWS of the tendons ex-vivo, buried in the couplant and placed in the isolated muscles showed significant differences (p < 0.001). The ex-vivo condition produced the highest SWS and CV values. There were significant differences in SWS of livers with different sizes placed in muscles (p < 0.001). The highest SWS value was associated with small pieces of livers. No significant difference was found in SWS acquired under different rectangular box sizes (p > 0.05).

CONCLUSIONS

Under the present study conditions, the surrounding environment of the target tissue makes a big difference to lesion SWS values. The lesion size will affect the assessment of its inherent elasticity. The size of the sampling frame has no significant effect on the tissue SWS.

摘要

目的

探讨目标组织周围环境、病变大小和矩形采样盒大小对剪切波速度(SWS)的影响。

方法

采用离体肌腱获取 SWS,然后将肌腱解剖并埋入偶联剂(凝胶)中,采用二维剪切波弹性成像(2D-SWE)进行评估。最后,将肌腱置于分离的肌肉中模拟肌内病变,并在两种矩形采样盒条件下测试其弹性。获取完整离体肝脏 SWS。同样,将肝脏的大块和小块取出,置于肌肉中,在两种矩形采样盒条件下用 SWE 进行评估。比较不同条件下获取的 SWS。采用变异系数(CV)评估变异性。采用组内相关系数(ICC)评估重复性。

结果

肌腱离体、埋入偶联剂和置于分离肌肉中的 SWS 有显著差异(p<0.001)。离体条件下产生的 SWS 和 CV 值最高。置于肌肉中的不同大小的肝脏的 SWS 有显著差异(p<0.001)。小块肝脏的 SWS 值最高。不同矩形盒大小下获得的 SWS 无显著差异(p>0.05)。

结论

在本研究条件下,目标组织的周围环境对病变 SWS 值有很大影响。病变大小会影响对其固有弹性的评估。采样框的大小对组织 SWS 无显著影响。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/d138/10113705/0408b23aa430/ACM2-24-e13924-g001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验