Qiu Chen, Situ Josephine, Wang Sheng-Ya, Vaghefi Ehsan
Department of Physiology, School of Medical Sciences, University of Auckland, New Zealand.
Department of Physics, University of Oxford, United Kingdom.
Biomed Opt Express. 2022 Oct 31;13(11):6136-6152. doi: 10.1364/BOE.468871. eCollection 2022 Nov 1.
Laser speckle contrast imaging (LSCI) can generate retinal blood flow maps inexpensively and non-invasively. These flow maps can be used to identify various eye disorders associated with reduced blood flow. Despite early success, one of the major obstacles to clinical adoption of LSCI is poor repeatability of the modality. Here, we propose an LSCI registration pipeline that registers contrast maps to correct for rigid movements. Post-registration, intra(same)-day and inter(next)-day repeatability are studied using various quantitative metrics. We have studied LSCI repeatability intra-day by using the coefficient of variation. Using the processing pipelines and custom hardware developed, similar repeatability was observed when compared to previously reported values in the literature. Inter-day repeatability analysis indicates no statistical evidence (p = 0.09) of a difference between flow measurements performed on two independent days. Further improvements to hardware, environmental controls, and participant control must be made to provide higher confidence in the repeatability of blood flow. However, this is the first time that repeatability across two different days (inter-day) using multiple exposure speckle imaging (MESI) has been analyzed and reported.
激光散斑对比成像(LSCI)能够以低成本、非侵入性的方式生成视网膜血流图。这些血流图可用于识别与血流减少相关的各种眼部疾病。尽管早期取得了成功,但LSCI临床应用的主要障碍之一是该方法的重复性较差。在此,我们提出了一种LSCI配准流程,该流程对对比图进行配准以校正刚性运动。配准后,使用各种定量指标研究日内(同一天)和日间(次日)重复性。我们通过使用变异系数研究了LSCI的日内重复性。使用所开发的处理流程和定制硬件,与文献中先前报道的值相比,观察到了相似的重复性。日间重复性分析表明,在两个独立日期进行的血流测量之间没有统计学证据(p = 0.09)表明存在差异。必须进一步改进硬件、环境控制和受试者控制,以提高对血流重复性的信心。然而,这是首次对使用多次曝光散斑成像(MESI)的两个不同日期(日间)的重复性进行分析和报道。