Department of Orthopedics, West China Hospital, Sichuan University, 37# Guoxue Lane, Chengdu Sichuan, 610041, China.
Department of Orthopedics, the First Affiliated Hospital of Zhengzhou University, Zhengzhou Henan, 450052, China.
BMC Musculoskelet Disord. 2023 Feb 11;24(1):115. doi: 10.1186/s12891-023-06215-1.
The bone-implant gap resulted from morphological mismatch between cervical bony endplates and implant footprint may have adverse impact on bone-implant interfacial osseointegration of cervical disc arthroplasty (CDA). The purpose of the study was to evaluate the impact of bone-implant gap size on the interfacial osseointegration in a rabbit animal model.
A series of round-plate implants with different teeth depth (0.5 mm, 1.0 mm, 1.5 mm and 2.0 mm) was specifically designed. A total of 48 New Zealand white rabbits were randomly categorized into four groups by the implants they received (0.5 mm: group A, 1.0 mm: group B, 1.5 mm: group C, 2.0 mm: group D). At 4 and 12 week after surgery, animals were sacrificed. Micro-CT, acid fuchsin and methylene blue staining and hematoxylin and eosin (HE) staining were conducted.
At 4 week and 12 week after surgery, both micro-CT and HE staining showed more new bone formation and larger bone coverage in group A and group B than that in group C and group D. At 12 week, the bone biometric parameters were significantly superior in group C when compared with group D (p < 0.05). At 12 week, hard tissue slicing demonstrated larger portion of direct contact of new bone to the HA coating in group A and group B.
Bone-implant gap size larger than 1.0 mm negatively affected bone-implant osseointegration between compact bone and HA coated implant surface.
颈椎骨终板与植入物足迹之间的形态不匹配导致的骨-植入物间隙可能对颈椎间盘置换术(CDA)的骨-植入物界面骨整合产生不利影响。本研究旨在评估兔动物模型中骨-植入物间隙大小对界面骨整合的影响。
专门设计了一系列具有不同齿深(0.5mm、1.0mm、1.5mm 和 2.0mm)的圆板植入物。将 48 只新西兰白兔随机分为四组,根据植入物分为 4 组(0.5mm:A 组、1.0mm:B 组、1.5mm:C 组、2.0mm:D 组)。手术后 4 周和 12 周处死动物。进行微 CT、酸性品红和亚甲蓝染色以及苏木精和伊红(HE)染色。
手术后 4 周和 12 周,微 CT 和 HE 染色均显示 A 组和 B 组的新骨形成和骨覆盖率均大于 C 组和 D 组。12 周时,C 组的骨生物计量学参数明显优于 D 组(p<0.05)。12 周时,硬组织切片显示 A 组和 B 组新骨与 HA 涂层的直接接触部分更大。
大于 1.0mm 的骨-植入物间隙大小会对密质骨和 HA 涂层植入物表面之间的骨-植入物骨整合产生负面影响。