He Tian, Pu Yin-Ying, Zhang Ya-Qin, Qian Zhe-Bin, Guo Le-Hang, Sun Li-Ping, Zhao Chong-Ke, Xu Hui-Xiong
Department of Medical Ultrasound, Center of Minimally Invasive Treatment for Tumor, Shanghai Tenth People's Hospital, Ultrasound Research and Education Institute, Clinical Research Center for Interventional Medicine, School of Medicine, Tongji University, Shanghai 200072, China.
Shanghai Engineering Research Center of Ultrasound Diagnosis and Treatment, Shanghai 200072, China.
Diagnostics (Basel). 2023 Jan 18;13(3):362. doi: 10.3390/diagnostics13030362.
Ultrasound (US) plays an important role in the diagnosis and management of breast diseases; however, effective breast US screening is lacking in rural and remote areas. To alleviate this issue, we prospectively evaluated the clinical availability of 5G-based telerobotic US technology for breast examinations in rural and remote areas.
Between September 2020 and March 2021, 63 patients underwent conventional and telerobotic US examinations in a rural island (Scenario A), while 20 patients underwent telerobotic US examination in a mobile car located in a remote county (Scenario B) in May 2021. The safety, duration, US image quality, consistency, and acceptability of the 5G-based telerobotic US were assessed.
In Scenario A, the average duration of the telerobotic US procedure was longer than that of conventional US (10.3 ± 3.3 min vs. 7.6 ± 3.0 min, = 0.017), but their average imaging scores were similar (4.86 vs. 4.90, = 0.159). Two cases of gynecomastia, one of lactation mastitis, and one of postoperative breast effusion were diagnosed and 32 nodules were detected using the two US methods. There was good interobserver agreement between the US features and BI-RADS categories of the identical nodules (ICC = 0.795-1.000). In Scenario B, breast nodules were detected in 65% of the patients using telerobotic US. Its average duration was 10.1 ± 2.3 min, and the average imaging score was 4.85. Overall, 90.4% of the patients were willing to choose telerobotic US in the future, and tele-sonologists were satisfied with 85.5% of the examinations.
The 5G-based telerobotic US system is feasible for providing effective breast examinations in rural and remote areas.
超声(US)在乳腺疾病的诊断和管理中发挥着重要作用;然而,农村和偏远地区缺乏有效的乳腺超声筛查。为缓解这一问题,我们前瞻性地评估了基于5G的远程机器人超声技术在农村和偏远地区进行乳腺检查的临床可用性。
2020年9月至2021年3月,63例患者在一个乡村岛屿接受了传统超声和远程机器人超声检查(场景A),而2021年5月,20例患者在一个偏远县的移动车内接受了远程机器人超声检查(场景B)。评估了基于5G的远程机器人超声的安全性、检查时长、超声图像质量、一致性和可接受性。
在场景A中,远程机器人超声检查的平均时长比传统超声长(10.3±3.3分钟对7.6±3.0分钟,P = 0.017),但它们的平均图像评分相似(4.86对4.90,P = 0.159)。使用两种超声方法诊断出2例男性乳腺增生、1例哺乳期乳腺炎和1例术后乳腺积液,并检测到32个结节。相同结节的超声特征和BI-RADS分类之间观察者间一致性良好(ICC = 0.795 - 1.000)。在场景B中,65%的患者使用远程机器人超声检测到乳腺结节。其平均时长为10.1±2.3分钟,平均图像评分为4.85。总体而言,90.4%的患者愿意在未来选择远程机器人超声,远程超声医生对85.5%的检查感到满意。
基于5G的远程机器人超声系统在农村和偏远地区提供有效的乳腺检查是可行的。