Department of Ergonomics, Faculty of Public Health, Iran University of Medical Sciences, Tehran, Iran.
Ambient Intelligence Laboratory (AMI-Lab), Computer Science Department, Faculty of Science, University of Sherbrooke, Sherbrooke, QC, Canada.
Work. 2023;76(1):403-410. doi: 10.3233/WOR-220185.
Laparoscopic surgeons who regularly perform endoscopy are more likely to develop musculoskeletal disorders than other internal medicine specialists, a difference that attributed to repetitive movements, poor postures, and sub-optimal equipment design.
This study aimed to design, build, and evaluate an endoscope holder for reducing the static load applied by the weight of the endoscope, in order to reduce musculoskeletal disorders risk factors in the surgeon's hand, shoulder and back issues regions.
A new endoscope holder was designed according to ergonomic design principles. The designed holder was evaluated by surface electromyography (sEMG) and discomfort assessment including 15 laparoscopic surgeons. The results were analyzed with centrality statistics and compared with the independent t-test using SPSS version 22.
The evaluation of the new endoscope holder shows a statistical significant decrease in the average electrical activity of biceps brachii, triceps brachii, lateral deltoid, T9 Thoracic erector spinae, L4 Lumbar erector spinae, and external oblique after using the holder (p < 0.05).
The results shows that using the new endoscope holder is associated with a lower level of discomfort, as well as a lower induced muscle activity. The results also highlight the need to upgrade the holder to offer rotability in all directions (perpendicular to the ground) which will be included in the next design.
与其他内科专家相比,经常进行内镜手术的腹腔镜外科医生更有可能患上肌肉骨骼疾病,这种差异归因于重复性动作、不良姿势和设备设计不理想。
本研究旨在设计、制造和评估一种内窥镜固定器,以减少内窥镜重量施加的静态负荷,从而降低外科医生手部、肩部和背部的肌肉骨骼疾病风险因素。
根据人体工程学设计原则设计了一种新的内窥镜固定器。使用表面肌电图(sEMG)和不适感评估对 15 名腹腔镜外科医生进行了评估。使用 SPSS 版本 22 对结果进行了中心性统计分析,并与独立 t 检验进行了比较。
新内窥镜固定器的评估显示,使用固定器后,肱二头肌、肱三头肌、三角肌外侧、T9 胸伸肌、L4 腰伸肌和外斜肌的平均电活动明显减少(p < 0.05)。
结果表明,使用新的内窥镜固定器与较低的不适感和较低的诱导肌肉活动相关。结果还强调需要升级固定器,以提供各个方向(垂直于地面)的可旋转性,这将包含在下一个设计中。