Talanta. 2023 May 15;257:124325. doi: 10.1016/j.talanta.2023.124325. Epub 2023 Feb 3.
Procalcitonin is a blood protein and precursor of the hormone calcitonin. The procalcitonin level increases due to bacterial infections, sepsis, and other related pathologies. Here, we present a simple biosensor for procalcitonin assay suitable for point-of-care tests as an alternative to the current laboratory methods. The biosensor was based on a QCM piezoelectric sensor and a conjugate of gold nanoparticles-antibodies conjugate. It was suitable for the procalcitonin assay in biological samples and fully correlated to the standard ELISA method, and it did not suffer false positive or negative results or interferences. The detection limit was equal to 37.8 ng/l and the quantification limit to 104 ng/l for a sample of 25 μl. The dynamic range of the assay was 37.8 ng/l to 30.0 μg/l. The practical relevance of the biosensor is expected considering the findings, and the possible application of the assay principle for the development of biosensors for other markers is inferred.
降钙素原是一种血液蛋白,也是降钙素的前体。降钙素原水平会因细菌感染、败血症和其他相关病理而升高。在这里,我们提出了一种用于降钙素原检测的简单生物传感器,适用于即时检测,可作为当前实验室方法的替代方法。该生物传感器基于 QCM 压电传感器和金纳米粒子-抗体结合物的共轭物。它适用于生物样品中的降钙素原检测,与标准 ELISA 方法完全相关,不会产生假阳性或假阴性结果或干扰。检测限为 37.8ng/l,定量限为 25μl 样品的 104ng/l。该测定的动态范围为 37.8ng/l 至 30.0μg/l。考虑到这些发现,预计该生物传感器具有实际意义,并且可以推断该测定原理可用于开发其他标志物的生物传感器。