Diagnosis (Berl). 2023 Feb 15;10(3):275-280. doi: 10.1515/dx-2022-0125. eCollection 2023 Aug 1.
Neurofilament light (NfL) chain is a marker of neuroaxonal damage in various neurological diseases. Here we quantitated NfL levels in the cerebrospinal fluid (CSF) and serum from patients with multiple sclerosis (MS) and controls, using the R-PLEX NfL assay, which employs advanced Meso Scale Discovery (MSD) electrochemiluminescence (ECL)-based detection technology.
NfL was quantitated in samples from 116 individuals from two sites (Ottawa Hospital Research Institute and Mayo Clinic), consisting of patients with MS (n=71) and age- and sex-matched inflammatory neurological controls (n=13) and non-inflammatory controls (n=32). Correlation of NfL levels between CSF and serum was assessed in paired samples in a subset of MS patients and controls (n=61). Additionally, we assessed the correlation between NfL levels obtained with MSD's R-PLEX and Quanterix's single molecule array (Simoa) assays in CSF and serum (n=32).
Using the R-PLEX, NfL was quantitated in 99% of the samples tested, and showed a broad range in the CSF (82-500,000 ng/L) and serum (8.84-2,014 ng/L). Nf-L levels in both biofluids correlated strongly (r=0.81, p<0.0001). Lastly, Nf-L measured by MSD's R-PLEX and Quanterix's Simoa assays were highly correlated for both biofluids (CSF: r=0.94, p<0.0001; serum: r=0.95, p<0.0001).
We show that MSD's R-PLEX NfL assay can reliably quantitate levels of NfL in the CSF and serum from patients with MS and controls, where levels correlate strongly with Simoa.
神经丝轻链(NfL)是各种神经疾病中神经轴突损伤的标志物。在这里,我们使用基于 MESO 规模发现(MSD)的电化学发光(ECL)检测技术的 R-PLEX NfL 测定法,定量了来自多发性硬化症(MS)患者和对照者的脑脊液(CSF)和血清中的 NfL 水平。
我们从两个地点(渥太华医院研究所和梅奥诊所)的 116 名个体中定量了 NfL,其中包括 MS 患者(n=71)、年龄和性别匹配的炎症性神经对照组(n=13)和非炎症性对照组(n=32)。在 MS 患者和对照组的亚组中(n=61),评估了 CSF 和血清中 NfL 水平的相关性。此外,我们评估了 MSD 的 R-PLEX 和 Quanterix 的单分子阵列(Simoa)测定法在 CSF 和血清中获得的 NfL 水平之间的相关性(n=32)。
使用 R-PLEX,在 99%的测试样本中定量了 NfL,CSF 中的范围很广(82-500,000ng/L),血清中的范围为(8.84-2,014ng/L)。两种生物流体中的 Nf-L 水平均强烈相关(r=0.81,p<0.0001)。最后,MSD 的 R-PLEX 和 Quanterix 的 Simoa 测定法测量的 Nf-L 在两种生物液中高度相关(CSF:r=0.94,p<0.0001;血清:r=0.95,p<0.0001)。
我们表明,MSD 的 R-PLEX NfL 测定法可以可靠地定量 MS 患者和对照者的 CSF 和血清中的 NfL 水平,其水平与 Simoa 密切相关。