Suppr超能文献

各种电子设备对电子根尖定位仪性能的影响

Effect of Various Electronic Devices on the Performance of Electronic Apex Locator.

作者信息

Parirokh Masoud, Manochehrifar Hamed, Abbott Paul V, Borna Roya, Haghdoost Ali Akbar

机构信息

Social Determinants on Oral Health Research Center, Kerman University of Medical Sciences, Kerman, Iran.

Endodontology Research Center, Kerman University of Medical Sciences, Kerman, Iran.

出版信息

Iran Endod J. 2019 Fall;14(4):278-282. doi: 10.22037/iej.v14i4.23752.

Abstract

INTRODUCTION

Apex locators are important devices that can be used to determine working length during root canal treatment. However, it is not known whether electronic devices with various wave length frequencies influence the accuracy of apex locators. The aim of this study was to determine whether a cordless phone set, MP4 player, FM radio, asymmetric digital subscriber line ADSL), or a mobile phone could influence the accuracy of working length determination by the Dentaport ZX apex locator.

METHODS AND MATERIALS

In this study, twenty-four sound intact single-rooted extracted human teeth were used. After determining the root canal length with a file and a microscope, the canals were measured with three separate Dentaport ZX apex locators while one of the test devices (ADSL, MP4, FM radio, mobile phone, and cordless phone) was used at a distance of 50 cm from the apex locator. Bland-Altman plots was used for reliability and consistency.

RESULTS

Except for the FM radio, all other devices showed significant difference with actual working length (<0.05). ADSL and MP4 player showed the least consistency compared to the other devices.

CONCLUSION

Based on this study, the use of ADSL, MP4 player, mobile phone, and cordless phone during root canal treatment may influence working length determination with the Dentaport ZX apex locator.

摘要

引言

根尖定位仪是根管治疗中用于确定工作长度的重要设备。然而,尚不清楚不同波长频率的电子设备是否会影响根尖定位仪的准确性。本研究的目的是确定无绳电话、MP4播放器、调频收音机、非对称数字用户线路(ADSL)或手机是否会影响Dentaport ZX根尖定位仪确定工作长度的准确性。

方法与材料

本研究使用了24颗完好无损的单根拔除人牙。在用锉和显微镜确定根管长度后,使用三个独立的Dentaport ZX根尖定位仪测量根管,同时将其中一个测试设备(ADSL、MP4、调频收音机、手机和无绳电话)放置在距根尖定位仪50厘米处。采用Bland-Altman图分析可靠性和一致性。

结果

除调频收音机外,所有其他设备与实际工作长度均有显著差异(<0.05)。与其他设备相比,ADSL和MP4播放器的一致性最差。

结论

基于本研究,根管治疗期间使用ADSL、MP4播放器、手机和无绳电话可能会影响Dentaport ZX根尖定位仪确定工作长度。