Comparison of in vivo and in vitro readings when testing the accuracy of the Root ZX apex locator.

机构信息

Department of Endodontics, Universitat Internacional de Catalunya, Barcelona, Spain.

出版信息

J Endod. 2012 Feb;38(2):236-9. doi: 10.1016/j.joen.2011.10.008. Epub 2011 Dec 5.

Abstract

INTRODUCTION

The purpose of the study was to compare the accuracy of the Root ZX electronic apex locator (J Morita Corp, Tokyo, Japan) between an in vivo and an in vitro model.

METHODS

The working length (WL) was determined electronically for 46 root canals of human teeth with a 15 K-file using both in vitro (n = 23) and in vivo (n = 23) models. The files were fixed at the WL. The apical 4 mm of each canal was trimmed to expose the file tip. The samples were observed under a scanning electron microscope, and the distance from the file tip to the point 0.5 mm coronal to the major foramen (the final WL) was measured. The data were analyzed using the Student t test, and significance was set at P ≤ .05.

RESULTS

The statistical analysis revealed no significant differences between the in vivo group and in vitro group with respect to the accuracy of the Root ZX device in determining the final WL. The mean distance from the final WL to the file tip was 0.23 ± 0.39 mm for the in vivo group and 0.29 ± 0.32 mm for the in vitro group. In determining the final WL, the Root ZX was accurate 78.3% of the time to ±0.5 mm and 100% of the time to ±1 mm in the in vivo group, whereas it was accurate 74% of the time to ±0.5 mm and 100% of the time to ±1 mm in the in vitro group.

CONCLUSIONS

No statistically significant differences were observed between the in vivo group and the in vitro group.

摘要

介绍

本研究的目的是比较 Root ZX 电子根尖定位仪(日本 Morita 株式会社)在体内和体外模型中的准确性。

方法

使用 15 K 锉对 46 个人类牙齿的根管进行电子根尖长度(WL)测量,分别采用体内(n=23)和体外(n=23)模型。将锉固定在 WL 处。每个根管的根尖 4mm 被修剪以暴露锉尖。将样本置于扫描电子显微镜下观察,测量从锉尖到距主孔 0.5mm 处的距离(最终 WL)。使用学生 t 检验对数据进行分析,P≤0.05 为差异有统计学意义。

结果

统计学分析显示,体内组和体外组在 Root ZX 设备确定最终 WL 的准确性方面无显著差异。体内组最终 WL 到锉尖的平均距离为 0.23±0.39mm,体外组为 0.29±0.32mm。在确定最终 WL 时,Root ZX 在体内组的准确率为 78.3%,在±0.5mm 以内,100%在±1mm 以内;在体外组的准确率为 74%,在±0.5mm 以内,100%在±1mm 以内。

结论

体内组和体外组之间未观察到统计学差异。