Cleaning and shaping of root canal system requires various irrigating solutions. The question is whether the presence of irrigation solution alters working length determination using apex locators. The purpose of this study was to compare the accuracy of Root ZX and i-Root apex locator for determining working length in the presence of different irrigating solution.
Eighty extracted single rooted human teeth were used. The teeth were sectioned at Cemento Enamel Junction (CEJ) and actual canal length determined. Then, working length measurements were obtained using Root ZX and i-Root apex locator in the presence of irrigating solutions namely 0.9% saline, 3% of sodium hypochlorite (NaOCl), 2% chlorhexidine digluconate(CHX) and 17% Ethylene Diamine Tetra Acetic acid (EDTA). The measurements obtained with Root ZX and i-Root apex locator were compared with actual canal length and subjected to statistical analysis using Post Hoc Test Tukey's Method.
This study revealed that both the tested Electronic apex locators (EAL) were able to measure the canal length in the presence of tested irrigating solutions. The presence of irrigating solutions of saline, NaOCl, chlorhexidine and EDTA in the root canal marginally influenced the accuracy of the Root ZX or i-Root (P < 0.36), but with no clinical significance.
Root ZX and i-Root can be used safely to determine working length in the presence of various irrigants. The content of the root canal did not influence the accuracy while measuring working length using Electronic apex locators (EAL).
根管系统的清理和塑形需要使用多种冲洗液。问题在于冲洗液的存在是否会改变使用根尖定位仪确定工作长度。本研究的目的是比较Root ZX和i-Root根尖定位仪在不同冲洗液存在情况下确定工作长度的准确性。
使用80颗拔除的单根人牙。在牙骨质釉质界(CEJ)处将牙齿截断并确定实际根管长度。然后,在存在冲洗液即0.9%生理盐水、3%次氯酸钠(NaOCl)、2%葡萄糖酸洗必泰(CHX)和17%乙二胺四乙酸(EDTA)的情况下,使用Root ZX和i-Root根尖定位仪获得工作长度测量值。将用Root ZX和i-Root根尖定位仪获得的测量值与实际根管长度进行比较,并使用事后检验Tukey方法进行统计分析。
本研究表明,两种受试电子根尖定位仪(EAL)在受试冲洗液存在的情况下均能够测量根管长度。根管中生理盐水、NaOCl、洗必泰和EDTA冲洗液的存在对Root ZX或i-Root的准确性有轻微影响(P < 0.36),但无临床意义。
Root ZX和i-Root可以安全地用于在存在各种冲洗剂的情况下确定工作长度。在使用电子根尖定位仪(EAL)测量工作长度时,根管内容物不会影响准确性。