Hwang Jin Ho, Park Sang Woo, Min Jeeyoung, Yang Woo Young, Kwon Yong Wonn, Hwang Jae Joon, Kim Jun Seok, Lee Song Am, Chee Hyun Keun
Department of Radiology, Konkuk University School of Medicine, Seoul, Korea.
Department of Radiology, Seoul Sun Vascular Pain Center, Seoul, Korea.
Ann Surg Treat Res. 2023 Mar;104(3):164-169. doi: 10.4174/astr.2023.104.3.164. Epub 2023 Feb 28.
This study aimed to determine the effectiveness and safety of a newly developed endovenous radiofrequency (RF) catheter compared with that of the existing RF catheter in a canine model.
Seven dogs underwent ablation using 1 control catheter (ClosureFAST, CF; Covidien) and 1 experimental catheter (VENISTAR, VS; STARmed Co., Ltd.) in the femoral and cephalic veins. The ablated vein was evaluated macroscopically (2,3,5-triphenyltetrazolium chloride staining, TTC), microscopically (hematoxylin and eosin staining), and ultrasonographically. Vessel injury score was used to evaluate the ablating effect objectively. Veins from 1 dog were evaluated on the day of ablation, while in the remaining 6 dogs, the ablated veins were evaluated 2 weeks later.
A total of 23 veins (CF, 11 veins; VS, 12 veins) were ablated in 7 dogs. Non-TTC-stained vein wall areas were identified in all ablated veins. No significant difference was observed in the mean vessel injury score (2.54 ± 1.16 2.42 ± 1.13, P = 0.656) and the mean vessel wall thickness (0.32 ± 0.03 mm 0.31 ± 0.05 mm, P = 0.212) between CF and VS. There was no blood flow in all veins ablated with VS, whereas there was remaining blood flow in 1 vein ablated with CF. Perivenous complication was not observed.
Endovenous RF ablation using a newly developed VS RF catheter seems to provide comparable occlusion rate and degree of vein wall injury without perivenous adverse events compared to the most commonly used RF catheter (CF).
本研究旨在确定在犬类模型中,新开发的静脉内射频(RF)导管与现有RF导管相比的有效性和安全性。
7只犬在股静脉和头静脉中使用1根对照导管(ClosureFAST,CF;柯惠医疗)和1根实验导管(VENISTAR,VS;STARmed有限公司)进行消融。对消融后的静脉进行宏观评估(2,3,5-三苯基氯化四氮唑染色,TTC)、微观评估(苏木精和伊红染色)以及超声评估。使用血管损伤评分客观评估消融效果。1只犬的静脉在消融当天进行评估,其余6只犬的消融静脉在2周后进行评估。
7只犬共消融了23条静脉(CF组11条静脉;VS组12条静脉)。在所有消融静脉中均发现了非TTC染色的静脉壁区域。CF组和VS组之间在平均血管损伤评分(2.54±1.16对2.42±1.13,P = 0.656)和平均血管壁厚度(0.32±0.03mm对0.31±0.05mm,P = 0.212)方面未观察到显著差异。VS消融的所有静脉均无血流,而CF消融的1条静脉仍有残留血流。未观察到静脉周围并发症。
与最常用的RF导管(CF)相比,使用新开发的VS RF导管进行静脉内RF消融似乎能提供相当的闭塞率和静脉壁损伤程度,且无静脉周围不良事件。