Suppr超能文献

有晶状体眼人工晶状体的拱顶预测分析:制造商计算器与理论公式与临床实践的比较。

Analysis of vault prediction in phakic implantable phakic collamer lenses: manufacturer's calculator vs theoretical formulae vs clinical practice.

机构信息

Faculty of Optics and Optometry, Complutense University, Madrid, Spain.

Clínica Rementería, 36 Almagro Street, 28010, Madrid, Spain.

出版信息

Graefes Arch Clin Exp Ophthalmol. 2023 Aug;261(8):2403-2409. doi: 10.1007/s00417-023-06016-1. Epub 2023 Mar 14.

Abstract

BACKGROUND

Implantable collamer phakic (ICL) intraocular lens sizing calculations are necessary to avoid complications associated to inadequate sizing. Historically, Holladay R, Dougherty, Hernández-Matamoros, and other authors have tried to create new formulas that solve calculation problems and provide higher reliability. In addition, in recent years, the appearance of new equipment, parameters, and formulas have led to significant progress. This paper compares the sizing according to manufacturer's method and other methods.

METHODS

Forty-three eyes of 24 patients with EVO ICL implanted, with at least 1 year of follow-up, were analysed. The analysed variables were white to white (WTW), anterior chamber depth (ACD), ACW (angle-to-angle), crystalline lens rise (CLR), ICL size, vault measured at 1 week and 1 year after surgery, ICL size, and vault predicted by Nakamura-2 as well as vault size predicted by Igarashi.

RESULTS

Sizing calculation with Online Calculation and Ordering System according to WTW and ACD is a good indicator with 86% success rate. The calculation with Nakamura 2 suggests larger ICL sizes in 32.5% of cases and smaller in 18.6% of cases, while the resulting Vault according to Igarashi obtains better results without significant differences.

CONCLUSIONS

ICL sizing according WTW and ACD, using the manufacturer's algorithm, seems to be the most predictable method compared to other algorithms using other variables. The surgeon's expertise also has a high importance in the final ICL size election.

摘要

背景

为避免因尺寸不合适而导致的并发症,有必要对植入性 Collamer 型有晶状体眼人工晶状体(ICL)进行尺寸计算。历史上,Holladay R、Dougherty、Hernández-Matamoros 和其他作者曾尝试创建新的公式来解决计算问题并提高可靠性。此外,近年来,新设备、参数和公式的出现带来了显著的进展。本文比较了根据制造商方法和其他方法进行的尺寸计算。

方法

分析了 24 名患者的 43 只眼,这些患者均植入了 EVO ICL,随访时间至少 1 年。分析的变量包括白到白(WTW)、前房深度(ACD)、角到角(ACW)、晶状体上升(CLR)、ICL 尺寸、术后 1 周和 1 年时的拱高测量值、ICL 尺寸、Nakamura-2 预测的拱高以及 Igarashi 预测的拱高尺寸。

结果

根据 WTW 和 ACD 使用在线计算和订购系统进行尺寸计算的成功率为 86%,是一个很好的指标。Nakamura 2 计算提示 32.5%的情况下需要更大的 ICL 尺寸,18.6%的情况下需要更小的尺寸,而根据 Igarashi 计算得到的拱高则获得了更好的结果,差异无统计学意义。

结论

与使用其他变量的其他算法相比,根据 WTW 和 ACD 使用制造商算法进行 ICL 尺寸计算似乎是最可预测的方法。术者的专业知识在最终 ICL 尺寸选择中也具有重要意义。