Evaluation of anastomosis quality with intraoperative transit time flowmeter in minimally invasive multi-vessel coronary artery bypass grafting via left anterior mini-thoracotomy.

作者信息

Çaynak Barış, Sicim Hüseyin

机构信息

Department of Cardiovascular Surgery, Private Medical Practice, Istanbul, Türkiye.

Department of Cardiovascular Surgery, Kırklareli Training and Research Hospital, Kırklareli, Türkiye.

出版信息

Turk Gogus Kalp Damar Cerrahisi Derg. 2023 Jan 30;31(1):56-62. doi: 10.5606/tgkdc.dergisi.2023.23744. eCollection 2023 Jan.

Abstract

BACKGROUND

In this study, we aimed to describe our intraoperative transit-time flow measurement results as an integral component of the operation and evaluate the graft patency and anastomosis quality in patients who underwent minimally invasive multi-vessel coronary artery bypass grafting via mini-thoracotomy.

METHODS

Between May 2020 and September 2021, a total of 45 consecutive patients (32 males, 13 females; mean age: 51.2±8.6 years; range, 41 to 72 years) who underwent minimally invasive multi-vessel coronary artery bypass grafting via left anterior mini-thoracotomy were retrospectively analyzed. We used the technique of intraoperative transit-time flowmetry in all patients. The patients were operated under cardiopulmonary bypass. A saphenous vein graft was used in all anastomoses, except for the left internal thoracic artery.

RESULTS

The mean left internal mammary artery flow rate was 36.2±14.1 mL/min, mean flow rate of the diagonal grafts was 48.2±13.1 mL/min, mean flow rate of the circumflex grafts was 41.2±21.1 mL/min, and mean flow rate of the right coronary artery grafts was 52.2±11.3 mL/min. Wave patterns and flow parameters of all grafts were normal in the intraoperative measurements, since the pulsatility index values in all anastomoses were within normal limits. The operation was completed after anastomotic openings and graft patency were ensured.

CONCLUSION

The use of an intraoperative flowmeter to show the graft patency and anastomosis quality gives confidence both to the surgeon and the patient. In multi-vessel coronary artery bypass grafting via mini-thoracotomy, anastomosis quality can be evaluated well with this technique.

摘要

背景

在本研究中,我们旨在描述术中通过经胸实时血流测定获得的结果,将其作为手术的一个组成部分,并评估接受经左前小切口微创多支冠状动脉旁路移植术患者的移植物通畅情况和吻合质量。

方法

回顾性分析2020年5月至2021年9月期间,连续45例接受经左前小切口微创多支冠状动脉旁路移植术的患者(男性32例,女性13例;平均年龄:51.2±8.6岁;范围41至72岁)。所有患者均采用术中经胸实时血流测定技术。患者在体外循环下进行手术。除左乳内动脉外,所有吻合均采用大隐静脉移植物。

结果

左乳内动脉平均血流速度为36.2±14.1 mL/min,对角支移植物平均血流速度为48.2±13.1 mL/min,回旋支移植物平均血流速度为41.2±21.1 mL/min,右冠状动脉移植物平均血流速度为52.2±11.3 mL/min。术中测量所有移植物的波形和血流参数均正常,因为所有吻合口的搏动指数值均在正常范围内。在确保吻合口开放和移植物通畅后完成手术。

结论

术中使用流量计显示移植物通畅情况和吻合质量,对外科医生和患者都具有信心。在经小切口进行多支冠状动脉旁路移植术中,该技术可以很好地评估吻合质量。