Transpalpebral intraocular pressure measurement by Diaton compared to Goldman applanation tonometer in myopic eyes before and after transepithelial photorefractive keratectomy in Saudi Arabia.

作者信息

Alzuhairy Sultan

机构信息

Department of Ophthalmology, College of Medicine, Qassim University, Buraidah 51452, Saudi Arabia.

出版信息

Int J Ophthalmol. 2023 Mar 18;16(3):375-381. doi: 10.18240/ijo.2023.03.07. eCollection 2023.

Abstract

AIM

To compare intraocular pressure (IOP) measured the trans palpebral IOP (tpIOP) method using a Diaton or with a Goldman applanation tonometer (GAT) and study the determinants of IOP difference in eyes undergoing transepithelial photorefractive keratectomy (TPRK) for myopia.

METHODS

This cross-sectional validation study was held in 2020 in an eye hospital in Saudi Arabia. IOP was measured by Diaton and GAT before treatment, after one week, and one month of TPRK. It is considered if IOP difference by Diaton and GAT was less than ±2 mm Hg acceptable. The IOP difference was tested if correlated to spherical equivalent (SE), central corneal thickness (CCT), age, gender, or tpIOP.

RESULTS

Totolly 200 myopic eyes of 100 patients were included in the study. The mean difference of IOP measured by two methods before TPRK, 1wk, and 1mo after TPRK were 0.790, -0.790, and -0.920 mm Hg, respectively (<0.001). Diaton could measure IOP effectively 89.5% eyes before TPRK and 82% and 84% at 1wk and 1mo after TPTK, respectively. At week 1, IOP differences were significantly correlated to baseline CCT (=0.02) and tpIOP at week 1 (<0.001). One month after TPRK, only tpIOP was significantly correlated to the difference in IOP (<0.001).

CONCLUSION

Diaton is a good screening tool for IOP before TPRK. It helps in monitoring IOP after surgery. Although more practical, it is less efficient than GAT. In eyes with high myopia and low tpIOP before surgery, IOP post-TPRK by Diaton and GAT could differ.

摘要

目的

比较使用Diaton经睑眼压测量法(tpIOP)与戈德曼压平眼压计(GAT)测量的眼压(IOP),并研究近视患者接受经上皮准分子激光角膜切削术(TPRK)后眼压差异的决定因素。

方法

这项横断面验证研究于2020年在沙特阿拉伯的一家眼科医院进行。在TPRK治疗前、治疗后1周和1个月,使用Diaton和GAT测量眼压。如果Diaton和GAT测量的眼压差异小于±2 mmHg,则认为是可接受的。测试眼压差异是否与等效球镜度(SE)、中央角膜厚度(CCT)、年龄、性别或tpIOP相关。

结果

本研究共纳入100例患者的200只近视眼睛。TPRK术前、术后1周和1个月两种方法测量的眼压平均差异分别为0.790、-0.790和-0.920 mmHg(<0.001)。Diaton在TPRK术前能有效测量89.5%的眼睛眼压,在TPRK术后1周和1个月分别能有效测量82%和84%的眼睛眼压。在第1周,眼压差异与基线CCT(=0.02)和第1周的tpIOP显著相关(<0.001)。TPRK术后1个月,只有tpIOP与眼压差异显著相关(<0.001)。

结论

Diaton是TPRK术前眼压的良好筛查工具。它有助于术后眼压监测。虽然更实用,但它比GAT效率低。术前高度近视且tpIOP低的眼睛,TPRK术后Diaton和GAT测量的眼压可能不同。

相似文献

7
Central corneal thickness and Diaton transpalpebral tonometry.中央角膜厚度与Diaton经睑眼压测量法
Graefes Arch Clin Exp Ophthalmol. 2008 Jun;246(6):881-9. doi: 10.1007/s00417-008-0769-8. Epub 2008 Feb 20.

本文引用的文献

4
Intraocular pressure measurement: A review.眼压测量:综述。
Surv Ophthalmol. 2022 Sep-Oct;67(5):1319-1331. doi: 10.1016/j.survophthal.2022.03.001. Epub 2022 Mar 4.
8
Refractive errors among Saudi college students and associated risk factors.沙特大学生的屈光不正及相关风险因素
Clin Ophthalmol. 2019 Feb 27;13:437-443. doi: 10.2147/OPTH.S193213. eCollection 2019.
10
Intraocular Pressure Fluctuation: Is It Important?眼压波动:它重要吗?
J Ophthalmic Vis Res. 2018 Apr-Jun;13(2):170-174. doi: 10.4103/jovr.jovr_35_18.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验