在肾上腺静脉采样前使用双能 CT 虚拟单能量图像改善肾上腺静脉可视化。

Improving the Visualization of the Adrenal Veins Using Virtual Monoenergetic Images from Dual-Energy Computed Tomography before Adrenal Venous Sampling.

机构信息

Department of Radiology, Sun Yat-sen Memorial Hospital, Sun Yat-sen University, No. 107 Yanjiang Road West, Guangzhou 510120, China.

Guangdong Provincial Key Laboratory of Malignant Tumor Epigenetics and Gene Regulation, Medical Research Center, Sun Yat-sen Memorial Hospital, Sun Yat-sen University, Guangzhou 510120, China.

出版信息

Tomography. 2023 Feb 23;9(2):485-496. doi: 10.3390/tomography9020040.

Abstract

(1) Background: This study explored the optimal energy level in advanced virtual monoenergetic images (VMI+) from dual-energy computed tomography angiography (DE-CTA) for adrenal veins visualization before adrenal venous sampling (AVS). (2) Methods: Thirty-nine patients were included in this prospective single-center study. The CT value, noise, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were measured in both adrenal veins and abdominal solid organs and were then compared between VMI+ within the range of 40-80 kiloelectron volt (keV). The visualization rate of the adrenal veins and the overall image quality of solid organs were subjectively compared among different keV VMI+. The AVS success rate was recorded for 20 patients. (3) Results: For the adrenal veins, 40 keV VMI+ had the peak CT value, noise and CNR ( < 0.05). Subjectively, the visualization rate was the highest at 40 keV (100% for the right adrenal vein, and 97.4% for the left adrenal vein) ( < 0.05). For solid organs, the CT value, noise and CNR at 50 keV were lower than those at 40 keV ( < 0.05), but the SNR was similar between 40 keV and 50 keV. The overall subjective image quality of solid organs at 50 keV was the best ( < 0.05). The AVS success rate was 95%. (4) Conclusions: For VMI+, 40 keV was the preferential energy level to obtain a high visualization rate of the adrenal veins and a high success rate of AVS, while 50 keV was the favorable energy level for the depiction of abdominal organs.

摘要

(1) 背景:本研究旨在探讨双能 CT 血管造影术(DE-CTA)衍生的高级虚拟单能量图像(VMI+)在肾上腺静脉取样(AVS)前显示肾上腺静脉的最佳能量水平。

(2) 方法:本前瞻性单中心研究纳入 39 例患者。分别测量肾上腺静脉和腹部实质脏器在 40-80keV 范围内的 CT 值、噪声、信噪比(SNR)和对比噪声比(CNR),并对其进行比较。比较不同 keV VMI+的肾上腺静脉显示率和实质脏器整体图像质量。记录 20 例患者的 AVS 成功率。

(3) 结果:对于肾上腺静脉,40keV VMI+的 CT 值、噪声和 CNR 最高(<0.05)。主观上,40keV 的显示率最高(右侧肾上腺静脉为 100%,左侧为 97.4%)(<0.05)。对于实质脏器,50keV 的 CT 值、噪声和 CNR 低于 40keV(<0.05),但 40keV 和 50keV 的 SNR 相似。50keV 时实质脏器的整体主观图像质量最佳(<0.05)。AVS 成功率为 95%。

(4) 结论:对于 VMI+,40keV 是获得高的肾上腺静脉显示率和高 AVS 成功率的首选能量水平,而 50keV 则有利于腹部脏器的显示。