BACKGROUND Rotary endodontic instruments are increasingly used in root canal treatment and have replaced stainless steel manual files. Cyclic fatigue is the cyclic loading of stress to produce deformation or fracture. This study aimed to evaluate and compare the effects of autoclave sterilization on cyclic fatigue in 5 types of rotary endodontic instruments. MATERIAL AND METHODS ProTaper Universal, BioRace, ProTaper Next, Twisted File, and HyFlex CM instruments were included in this study. Each type included 96 instruments, divided into 4 groups according to the number of sterilization cycles (0, 1, 3, 5). After sterilization, each group of instruments was divided into 2 subgroups and tested for cyclic fatigue in 2 simulated canals (45 degrees both and 2 radii, 2 mm and 5 mm). The number of cycles to failure (NCF) was calculated, and statistical analyses were carried out using the t test, Mann-Whitney U test, and ANOVA, followed by the Tukey post hoc test (p<0.05). Fracture surfaces were analyzed using scanning electron microscopy (SEM). RESULTS Within the group of non-sterilized instruments, ProTaper Universal showed significantly lower resistance to cyclic fatigue compared to the other types of instruments (p<0.001). After repeated sterilization, a significantly higher mean of NCF was observed for BioRace (p<0.001), ProTaper Next (p<0.001), Twisted File (p<0.001), and HyFlex CM (p<0.001) compared to ProTaper Universal. The resistance of HyFlex CM was significantly higher compared to the other types of instruments (p<0.001). CONCLUSIONS The findings from this study showed that autoclave sterilization of newer rotary endodontic instruments could increase resistance to cyclic fatigue.
旋转根管器械在根管治疗中越来越多地被使用,并且已经取代了不锈钢手动锉。循环疲劳是指循环加载应力以产生变形或断裂。本研究旨在评估和比较 5 种不同类型的旋转根管器械经高压蒸汽灭菌对循环疲劳的影响。
本研究纳入 ProTaper Universal、BioRace、ProTaper Next、Twisted File 和 HyFlex CM 这 5 种根管锉。每种类型包括 96 个器械,根据灭菌循环次数(0、1、3、5)分为 4 组。灭菌后,每组器械分为 2 个亚组,在 2 个模拟根管中进行循环疲劳测试(45 度和 2 个半径,2mm 和 5mm)。计算失效循环数(NCF),并采用 t 检验、Mann-Whitney U 检验和 ANOVA 进行统计学分析,随后采用 Tukey 事后检验(p<0.05)。使用扫描电子显微镜(SEM)分析断裂面。
在未经灭菌的器械组中,ProTaper Universal 与其他类型的器械相比,循环疲劳阻力明显较低(p<0.001)。经多次灭菌后,BioRace(p<0.001)、ProTaper Next(p<0.001)、Twisted File(p<0.001)和 HyFlex CM(p<0.001)的平均 NCF 值明显高于 ProTaper Universal。与其他类型的器械相比,HyFlex CM 的阻力明显更高(p<0.001)。
本研究结果表明,新型旋转根管器械经高压蒸汽灭菌可提高其对循环疲劳的抵抗力。