Aspide Raffaele, Bertolini Giacomo, Belotti Laura Maria Beatrice, Albini Riccioli Luca, Toni Francesco, Mazzatenta Diego, Palandri Giorgio, Vetrugno Luigi, Biasucci Daniele Guerino
IRCCS Istituto delle Scienze Neurologiche di Bologna, Anesthesia and Neurointensive Care Unit, 40139 Bologna, Italy.
Department of Biomedical and Neuromotor Sciences (DIBINEM), University of Bologna, 40126 Bologna, Italy.
J Clin Med. 2023 Apr 5;12(7):2713. doi: 10.3390/jcm12072713.
The measurement of optic nerve sheath diameter (ONSD) as a non-invasive method of estimating intracranial pressure has been widely reported in the literature. However, few studies have evaluated the accuracy of magnetic resonance imaging (MRI) in assessing ONSD measurements, although it is considered a very reliable method, it is not easily repeatable, expensive and is not readily available bedside. Herein, an assessment of the intra- and inter-rater reliability of ONSD assessment using MRI was conducted.
A consecutive, prospective cohort of patients with suspected idiopathic normal-pressure hydrocephalus was analyzed. ONSD MRI measurements of the transverse and sagittal diameters at a distance of 3 mm behind the papilla were evaluated twice each by two expert neuroradiologists. The correlations between MRI examiners were calculated using the concordance correlation coefficient (CCC).
Fifty patients were included in the study. ONSD MRI average measurements were substantially higher than clinically expected (>5 mm). Considering intra-rater concordance, only one of the two neuroradiologists achieved an excellent score at CCC. Only a moderate inter-observer CCC for MRI assessment was found at all diameters.
The use of a widespread MRI sequence (3D T1) to measure ONSD is not an accurate method since it may overestimate measurements and is dependent upon an operator.
作为一种估计颅内压的非侵入性方法,视神经鞘直径(ONSD)的测量在文献中已有广泛报道。然而,尽管磁共振成像(MRI)被认为是一种非常可靠的方法,但很少有研究评估其在评估ONSD测量中的准确性,因为它不易重复、昂贵且床边不易获得。在此,我们对使用MRI评估ONSD的评分者内和评分者间可靠性进行了评估。
对疑似特发性正常压力脑积水的连续前瞻性队列患者进行分析。两名神经放射学专家对乳头后3mm处的横径和矢状径进行两次ONSD MRI测量评估。使用一致性相关系数(CCC)计算MRI检查者之间的相关性。
50名患者纳入研究。ONSD MRI平均测量值显著高于临床预期(>5mm)。考虑评分者内一致性,两名神经放射学专家中只有一人在CCC上获得了优异分数。在所有直径上,MRI评估的观察者间CCC仅为中等。
使用广泛的MRI序列(3D T1)测量ONSD不是一种准确的方法,因为它可能高估测量值且依赖于操作者。