Screening for urinary tract infection using Bac-T-Screen bacteriuria device.

作者信息

Khanna O P, Son D L

出版信息

Urology. 1986 May;27(5):424-8. doi: 10.1016/0090-4295(86)90407-3.

Abstract

The Bac-T-Screen was used to process 795 urine specimens. Tests for urine specimens took slightly more than two minutes. The Bac-T-Screen predicted with 99 per cent accuracy if a specimen was negative for bacteriuria or pyuria. Thus, approximately one half of all carefully collected urine specimens need no further laboratory culture. In addition, the Bac-T-Screen detected bacteriuria with a sensitivity of 96 per cent at the 10(5) CFU/ml level of probability. Pyuria (1 +) was detected with a sensitivity of 98 per cent. The Bac-T-Screen can be used in an office practice as well as in the clinical laboratory.