Suppr超能文献

软件更新对口腔内扫描仪准确性的影响。

The impact of software updates on accuracy of intraoral scanners.

机构信息

Department of Prosthodontics, Semmelweis University, Szentkiralyi Street 47, 1088, Budapest, Hungary.

出版信息

BMC Oral Health. 2023 Apr 15;23(1):219. doi: 10.1186/s12903-023-02926-y.

Abstract

BACKGROUND

Digital workflow is showing an increasing tendency in everyday dentistry. Accuracy is essential during digital dental workflows for all indication areas. The present study aimed to evaluate the effect of software updates on the accuracy of intraoral scanner (IOS) devices.

METHODS

3Shape Trios 3 Pod with software versions 18.1.2. (TRI3_1) and 20.1.2. (TRI3_2); 3Shape Trios 4 Move, version 19.2.2. (TRI4_1); and 3Shape Trios 4 Pod, version 20.1.1. (TRI4_2) were used to take direct optical impressions from a polymethyl methacrylate (PMMA) full arch reference model with prepared teeth (FDI 11,14,17 for crowns and FDI 26 for onlay) and an edentulous region (between FDI 14 and 17). The scanners were used eight times; STL files were imported into Geomagic Control X for accuracy assessment by comparing them to a reference data set created by an industrial high-precision scanner (AICON SmartScan-3D C5). The average deviation of the surface points was calculated in three locations: across a full arch (Parameter 1), the region of a four-unit bridge (Parameter 2), and a single prepared abutment (Parameter 3).

RESULTS

In parameter 1 and 2, the newest model with the latest software (TRI4_2) reached the highest accuracy (31.06 ± 5.24 µm and 21.69 ± 7.50 µm). In parameter 3, an older generation scanner running legacy software produced the highest accuracy: TRI4_1, 11.75 ± 0.35 µm.

CONCLUSION

Appropriate software updates can significantly increase the trueness and precision of intraoral scanner devices. With updated software, the older generation can match the accuracy level of latest equipment.

摘要

背景

数字化工作流程在日常牙科中呈现出日益增长的趋势。在所有适应证领域,数字化牙科工作流程中准确性至关重要。本研究旨在评估软件更新对口腔内扫描仪(IOS)设备准确性的影响。

方法

使用 3Shape Trios 3 Pod,软件版本 18.1.2(TRI3_1)和 20.1.2(TRI3_2);3Shape Trios 4 Move,版本 19.2.2(TRI4_1);以及 3Shape Trios 4 Pod,版本 20.1.1(TRI4_2),从带有预备牙的聚甲基丙烯酸甲酯(PMMA)全牙弓参考模型(牙冠 FDI 11、14、17,嵌体 FDI 26)和无牙区(FDI 14 和 17 之间)获取直接光学印模。扫描仪共使用了 8 次;将 STL 文件导入 Geomagic Control X 中,通过与工业高精度扫描仪(AICON SmartScan-3D C5)创建的参考数据集进行比较来评估准确性。在三个位置计算表面点的平均偏差:全牙弓(参数 1)、四单位桥区域(参数 2)和单个预备基牙(参数 3)。

结果

在参数 1 和 2 中,最新模型搭配最新软件(TRI4_2)达到最高精度(31.06±5.24 µm 和 21.69±7.50 µm)。在参数 3 中,运行旧版软件的较旧一代扫描仪产生了最高的精度:TRI4_1,11.75±0.35 µm。

结论

适当的软件更新可以显著提高口腔内扫描仪设备的准确性和精密度。使用更新后的软件,旧一代设备可以达到最新设备的精度水平。