Abstract

OBJECTIVE

The aim of this study was to perform and evaluate PET/computed tomography acceptance tests separately using American Association of Physicists in Medicine (AAPM) Task Group 126 and National Electrical Manufacturers Association (NEMA) methods.

MATERIALS AND METHODS

Measurements of sensitivity, spatial resolution, count rate performance and scatter fraction, the accuracy of corrections for count losses and randoms, and image quality were obtained according to NEMA NU-2018. Likewise, the performance tests were made using the AAPM Task Group 126 method, and the results were compared with NEMA NU-2018.

RESULTS

The sensitivity at the isocenter was 8.87 cps/kBq according to NEMA and 7.60 cps/kBq by using the AAPM Task Group 126. For the spatial resolution, the full width at half maximum (FWHM) and FWTM values were 4.34 mm and 6.78 mm at 1 cm radial offset by NEMA, while AAPM Task Group 126 yielded FWHM and FWTM values of 4.42 mm and 8.14 mm, respectively. In the image quality, NEMA exhibited hot lesions contrast of 40.8, 56.7, 69.9, and 77.3 for 10, 13, 17, and 22 mm spheres, respectively. As a ratio to 25 mm, the "Hot" max standard uptake values by AAPM Task Group 126 were found to be 1, 1.1, 1.37, and 1.68 for 8, 12, 16, and 25 mm lesions, respectively.

CONCLUSION

Acceptance tests using NEMA are of high relevance and convenience for the reliability of the results. Alternatively, AAPM Task Group 126 seems convenient and more economical to apply with reliable outcomes for the equivalent tests.

摘要

目的

本研究旨在分别使用美国医学物理学家协会(AAPM)工作组 126 和美国电器制造商协会(NEMA)方法进行 PET/计算机断层扫描(CT)的验收测试,并对其进行评估。

材料与方法

根据 NEMA NU-2018,对灵敏度、空间分辨率、计数率性能和散射分数、计数损失和随机计数校正的准确性以及图像质量进行了测量。同样,使用 AAPM 工作组 126 方法进行了性能测试,并将结果与 NEMA NU-2018 进行了比较。

结果

根据 NEMA,在等中心处的灵敏度为 8.87 cps/kBq,而使用 AAPM 工作组 126 则为 7.60 cps/kBq。对于空间分辨率,NEMA 在 1cm 径向偏移时的全宽半最大值(FWHM)和 FWTM 值分别为 4.34mm 和 6.78mm,而 AAPM 工作组 126 的 FWHM 和 FWTM 值分别为 4.42mm 和 8.14mm。在图像质量方面,NEMA 分别显示 10、13、17 和 22mm 球体的热病变对比度为 40.8、56.7、69.9 和 77.3。与 25mm 相比,AAPM 工作组 126 的“热”最大标准摄取值分别为 1、1.1、1.37 和 1.68,用于 8、12、16 和 25mm 病变。

结论

使用 NEMA 的验收测试对于结果的可靠性具有重要意义和方便性。或者,AAPM 工作组 126 似乎更方便且更经济实惠,对于等效测试,可获得可靠的结果。