Near-infrared (NIR) fluorescence-guided surgery is increasingly utilized in humans and pets. As clinical imaging systems are optimized for Indocyanine green (ICG) detection, the usage of targeted dyes necessitates the validation of these systems for each dye. We investigated the impact of skin pigmentation and tissue overlay on the sensitivity of two NIR cameras (IC-Flow, Visionsense VS3 Iridum) for the detection of non-targeted (ICG, IRDye800) and targeted (Angiostamp, FAP-Cyan) NIR fluorophores in an big animal model.
We quantitatively measured the limit of detection (LOD) and signal-to-background ratio (SBR) and implemented a semi-quantitative visual score to account for subjective interpretation of images by the surgeon.
Visionsense VS3 Iridum outperformed IC-Flow in terms of LOD and SBR for the detection of all dyes except FAP-Cyan. Median SBR was negatively affected by skin pigmentation and tissue overlay with both camera systems. Level of agreement between quantitative and semi-quantitative visual score and interobserver agreement were better with Visionsense VS3 Iridum.
The overlay of different tissue types and skin pigmentation may negatively affect the ability of the two tested camera systems to identify nanomolar concentrations of targeted-fluorescent dyes and should be considered when planning surgical applications.
近红外(NIR)荧光引导手术在人类和宠物中的应用日益广泛。由于临床成像系统是针对吲哚菁绿(ICG)检测进行优化的,因此使用靶向染料需要针对每种染料对这些系统进行验证。我们在一个大型动物模型中研究了皮肤色素沉着和组织覆盖对两台近红外相机(IC-Flow、Visionsense VS3 Iridum)检测非靶向(ICG、IRDye800)和靶向(血管标记物、FAP-青色)近红外荧光团灵敏度的影响。
我们定量测量了检测限(LOD)和信号背景比(SBR),并实施了半定量视觉评分,以考虑外科医生对图像的主观解读。
在检测除FAP-青色以外的所有染料时,Visionsense VS3 Iridum在LOD和SBR方面均优于IC-Flow。两种相机系统的中位SBR均受到皮肤色素沉着和组织覆盖的负面影响。Visionsense VS3 Iridum在定量和半定量视觉评分之间的一致性水平以及观察者间一致性方面表现更好。
不同组织类型的覆盖和皮肤色素沉着可能会对两种测试相机系统识别纳摩尔浓度靶向荧光染料的能力产生负面影响,在规划手术应用时应予以考虑。