Lasers Med Sci. 2023 May 12;38(1):123. doi: 10.1007/s10103-023-03779-1.
The transient apical pressure side effect is an important safety consideration for Er:YAG laser-activated irrigation (LAI). Therefore, this study aimed to measure the transient apical peak pressure (TAPP) of LAI under different laser settings in various tooth models using a high-frequency sensor system. Tooth models with different pulp chamber structures, apical diameters, and curvatures were prepared using transparent resin and filled with deionised water. The Er:YAG laser fibre was placed 3 mm from the root canal orifice. Irrigation was performed at 10-40 mJ and 20-50 Hz using the super short pulse mode. The TAPP was measured using a 50,000-sample/second pressure sensor connected to the models' apices. The TAPP of LAI was significantly higher than that of other chemical preparation methods. Among all investigated factors, pulp chamber anatomy and apical diameters had the greatest effects and were highly related to the apical peak pressure. Root canal curvature showed no direct correlation with TAPP. The larger the final prepared working width, the greater the TAPP. Furthermore, both pulse energy and frequency had positive correlations with TAPP. In conclusion, tooth anatomy factors and laser parameter settings influenced TAPP during Er:YAG LAI. Therefore, proper settings of laser parameters are important to improve the safety of Er:YAG LAI.
激光激活冲洗时的暂短顶端压效应是 Er:YAG 激光的一个重要安全考虑因素。因此,本研究旨在使用高频压力传感器系统,针对不同的牙齿模型和不同的激光设置,测量激光激活冲洗时的暂短顶端峰值压力(TAPP)。使用透明树脂制备具有不同牙髓腔结构、根尖直径和曲率的牙齿模型,并在其中填充去离子水。将 Er:YAG 激光光纤放置在距根管口 3mm 处。使用超短脉冲模式,以 10-40mJ 和 20-50Hz 的频率进行冲洗。通过与模型根尖相连的 50000 样本/秒压力传感器测量 TAPP。激光激活冲洗时的 TAPP 明显高于其他化学预备方法。在所研究的所有因素中,牙髓腔解剖结构和根尖直径的影响最大,与根尖峰值压力高度相关。根管弯曲度与 TAPP 无直接相关性。最终预备的工作宽度越大,TAPP 越大。此外,脉冲能量和频率与 TAPP 均呈正相关。总之,牙齿解剖因素和激光参数设置会影响 Er:YAG 激光激活冲洗时的 TAPP。因此,适当设置激光参数对于提高 Er:YAG 激光激活冲洗的安全性非常重要。