Performance of the FreeStyle Libre Flash Glucose Monitoring System during an Oral Glucose Tolerance Test and Exercise in Healthy Adolescents.

机构信息

School of Sport, Exercise and Health Sciences, Loughborough University, Loughborough LE11 3TU, UK.

Clinical Nutrition Department, Faculty of Applied Medical Sciences, King Abdulaziz University, Jeddah 21589, Saudi Arabia.

出版信息

Sensors (Basel). 2023 Apr 25;23(9):4249. doi: 10.3390/s23094249.

Abstract

This study's aim was to assess FreeStyle Libre Flash glucose monitoring (FGM) performance during an oral glucose tolerance test (OGTT) and treadmill exercise in healthy adolescents. This should advance the feasibility and utility of user-friendly technologies for metabolic assessments in adolescents. Seventeen healthy adolescents (nine girls aged 12.8 ± 0.9 years) performed an OGTT and submaximal and maximal treadmill exercise tests in a laboratory setting. The scanned interstitial fluid glucose concentration ([ISFG]) obtained by FGM was compared against finger-prick capillary plasma glucose concentration ([CPG]) at 0 (pre-OGTT), -15, -30, -60, -120 min post-OGTT, pre-, mid-, post- submaximal exercise, and pre- and post- maximal exercise. Overall mean absolute relative difference (MARD) was 13.1 ± 8.5%, and 68% ( = 113) of the paired glucose data met the ISO 15197:2013 criteria. For clinical accuracy, 84% and 16% of FGM readings were within zones A and B in the Consensus Error Grid (CEG), respectively, which met the ISO 15197:2013 criteria of having at least 99% of results within these zones. Scanned [ISFG] were statistically lower than [CPG] at 15 (-1.16 mmol∙L, < 0.001) and 30 min (-0.74 mmol∙L, = 0.041) post-OGTT. Yet, post-OGTT glycaemic responses assessed by total and incremental areas under the curve (AUCs) were not significantly different, with trivial to small effect sizes ( ≥ 0.084, d = 0.14-0.45). Further, [ISFGs] were not different from [CPGs] during submaximal and maximal exercise tests (interaction ≥ 0.614). FGM can be a feasible alternative to reflect postprandial glycaemia (AUCs) in healthy adolescents who may not endure repeated finger pricks.

摘要

本研究旨在评估在健康青少年进行口服葡萄糖耐量试验(OGTT)和跑步机运动期间 FreeStyle Libre Flash 血糖监测(FGM)的性能。这应该推进在青少年代谢评估中使用用户友好型技术的可行性和实用性。17 名健康青少年(9 名女孩,年龄 12.8 ± 0.9 岁)在实验室环境中进行了 OGTT 和亚最大和最大跑步机运动测试。FGM 获得的扫描间质液葡萄糖浓度 ([ISFG]) 与指尖毛细血管血浆葡萄糖浓度 ([CPG]) 进行比较,时间点为 OGTT 前(-15、-30、-60、-120 分钟)、亚最大运动前、中、后和最大运动前、后。总体平均绝对相对差异(MARD)为 13.1 ± 8.5%,68%(=113)的配对血糖数据符合 ISO 15197:2013 标准。对于临床准确性,FGM 读数中有 84%和 16%分别落在共识误差网格(CEG)的 A 区和 B 区,符合 ISO 15197:2013 标准,至少 99%的结果在这些区域内。OGTT 后 15 分钟(-1.16 mmol·L,<0.001)和 30 分钟(-0.74 mmol·L,=0.041)时,扫描的 [ISFG] 明显低于 [CPG]。然而,OGTT 后总血糖和增量曲线下面积(AUCs)评估的血糖反应没有显著差异,效应大小为小到中等(≥0.084,d=0.14-0.45)。此外,在亚最大和最大运动测试期间,[ISFGs]与 [CPGs] 没有差异(交互作用≥0.614)。FGM 可以成为反映可能无法忍受反复指尖采血的健康青少年餐后血糖(AUCs)的可行替代方法。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验