Mansoori Tarannum, Charan Aknoor Sree Ram, Suresh Narala, Pesala Veerendranath, Nagalla Balakrishna
Department of Glaucoma, Anand Eye Institute, Hyderabad, Telangana, India.
Department of Retina, Anand Eye Institute, Hyderabad, Telangana, India.
J Curr Ophthalmol. 2023 Apr 29;34(4):436-441. doi: 10.4103/joco.joco_168_22. eCollection 2022 Oct-Dec.
To determine the intrasession, intraobserver, interobserver, and repeatability of choroidal thickness measurements in the healthy subjects imaged on enhanced depth imaging system of RTVue XR spectral domain optical coherence tomography (OCT).
In this prospective, cross-sectional study, seventy eyes of 70 healthy volunteers with no known ocular disease were imaged using high-density scanning protocol of RTVue XR OCT. In a single imaging session, three sequential 12 mm macular-enhanced depth horizontal line scans were obtained through the fovea. Two experienced examiners measured the subfoveal choroidal thickness (SFCT), choroidal thickness at 500 μm nasally and temporally from the fovea in all the eyes, using the manual calipers provided in the software. The graders were masked to each other's measurement readings. The coefficient of repeatability (CR) and intraclass correlation coefficient (ICC) were used to measure the reliability within graders. Intergrader variability was assessed using Bland-Altman method and 95% limits of agreement (LoA).
Intragrader CR for grader one was 4.11 μm (95% confidence interval [CI], -2.84-11.06) for SFCT and 5.73 μm (95% CI, -3.71-15.16) for the grader two. Intragrader ICC of grader one ranged from 0.996 for SFCT to 0.994 for temporal choroidal thickness. Intragrader ICC of grader two ranged from 0.993 for temporal choroidal thickness to 0.991 for SFCT. Intergrader CR ranged from 5.24 μm (95% CI, -4.66-15.15) for SFCT to 5.89 μm (95% CI, -7.27-19.04) for temporal choroidal thickness. Intergrader 95% LoA for SFCT, nasal and temporal choroidal thickness were, -15.84-12.15 μm, -15.99-17.7 μm, and - 19.12-15.57 μm, respectively.
Choroidal thickness measurements can be quantified with good repeatability using RTVue XR OCT, which would be useful in patients with chorioretinal diseases.
确定在使用RTVue XR光谱域光学相干断层扫描(OCT)增强深度成像系统成像的健康受试者中,脉络膜厚度测量的扫描期间、观察者内、观察者间及可重复性。
在这项前瞻性横断面研究中,对70名无已知眼部疾病的健康志愿者的70只眼,使用RTVue XR OCT的高密度扫描协议进行成像。在单次成像过程中,通过中央凹获得三条连续的12毫米黄斑增强深度水平线扫描。两名经验丰富的检查者使用软件中提供的手动卡尺,测量所有眼睛的中央凹下脉络膜厚度(SFCT)、距中央凹鼻侧和颞侧500μm处的脉络膜厚度。分级者对彼此的测量读数不知情。使用重复性系数(CR)和组内相关系数(ICC)来测量分级者内部的可靠性。使用Bland-Altman方法和95%一致性界限(LoA)评估分级者间的变异性。
分级者一的SFCT观察者内CR为4.11μm(95%置信区间[CI],-2.84-11.06),分级者二为5.73μm(95%CI,-3.71-15.16)。分级者一的观察者内ICC范围从SFCT的0.996到颞侧脉络膜厚度的0.994。分级者二的观察者内ICC范围从颞侧脉络膜厚度的0.993到SFCT的0.991。分级者间CR范围从SFCT的5.24μm(95%CI,-4.66-15.15)到颞侧脉络膜厚度的5.89μm(95%CI,-7.27-19.04)。SFCT、鼻侧和颞侧脉络膜厚度的分级者间95%LoA分别为-15.84-12.15μm、-15.99-17.7μm和-19.12-15.57μm。
使用RTVue XR OCT可以很好地重复量化脉络膜厚度测量,这对脉络膜视网膜疾病患者有用。