Suppr超能文献

3D 环型 CZT 系统,具有全新的低能和中能范围:超快双同位素肺 SPECT/CT 改善。

3D-Ring CZT System With New Low- and Medium-Energy Range: Ultrafast Dual-Isotope Lung SPECT/CT Improvement.

机构信息

From the Nuclear Medicine Department, CHU ORLEANS, Orleans, France.

出版信息

Clin Nucl Med. 2023 Aug 1;48(8):716-718. doi: 10.1097/RLU.0000000000004706. Epub 2023 May 24.

Abstract

New 3D-ring CZT systems with low- and medium-energy-range detectors allow for simultaneous dual-isotope lung scintigraphy. We compared 10-, 7-, 5-, and 3-minute acquisitions for 99m Tc and 81m Kr performed simultaneously on StarGuide CZT-SPECT/CT and reframed in 50 patients. Ventilation/perfusion mismatches were calculated (mean, 15.6% ± 28%), and Spearman correlation coefficients of mismatches were 0.994, 0.994, and 0.984 between 10- and 7-, 5-, and 3-minute acquisitions, respectively. No visual difference in image quality or final diagnosis was found. 3D-ring CZT-SPECT with low and medium energy range detectors allows ultrafast dual-isotope lung scintigraphy up to 3 minutes.

摘要

新型三维环 CZT 系统与低能和中能探测器相结合,可实现同时进行双同位素肺闪烁显像。我们比较了在 StarGuide CZT-SPECT/CT 上同时进行的 99mTc 和 81mKr10 分钟、7 分钟、5 分钟和 3 分钟采集,并在 50 例患者中重新进行了 50 分钟的采集。计算了通气/灌注不匹配(平均值为 15.6%±28%),10-分钟和 7-分钟、5-分钟和 3-分钟采集之间的不匹配的斯皮尔曼相关系数分别为 0.994、0.994 和 0.984。在图像质量或最终诊断方面没有发现明显差异。低能和中能探测器的三维环 CZT-SPECT 可实现超快的双同位素肺闪烁显像,最短采集时间可达 3 分钟。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/2cef/10317294/ee90f86fd167/cnm-48-716-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验