To evaluate the reliability and efficiency of sterile corneal thickness measurements of donor corneas stored in a plastic culture flask filled with organ culture medium I (MI) or II (MII) based on tomographic data using two different software: the built-in software of the anterior segment OCT (AS-OCT) and a MATLAB self-programmed software.
Twenty-five (25) donor corneas (50%) stored in MI and 25 (50%) in MII were imaged 5 times consecutively using an AS-OCT. The central corneal thickness (CCT) was measured both with the manual measurement tool of the AS-OCT (=CCTm) and with a MATLAB self-programmed software allowing (semi-)automated analysis (=CCTa). We analyzed the reliability of CCTm and CCTa using Cronbach´s alpha (α) and Wilcoxon signed-rank test.
Concerning CCTm, 68 measurements (54.4%) in MI and 46 (36.8%) in MII presented distortions in the imaged 3D-volumes and were discarded. Concerning CCTa, 5 (4%) in MI and 1 (0.8%) in MII were not analyzable. The mean (± SD) CCTm was 1129 ± 6.8 in MI and 820 ± 5.1 µm in MII. The mean CCTa was 1149 ± 2.7 and 811 ± 2.4 µm, respectively. Both methods showed a high reliability with a Cronbach´s α for CCTm of 1.0 (MI/MII) and for CCTa of 0.99 (MI) and 1.0 (MII). Nevertheless, the mean SD of the 5 measurements was significantly higher for CCTm compared to CCTa in MI (p = 0.03), but not in MII (p = 0.92).
Sterile donor tomography proves to be highly reliable for assessment of CCT with both methods. However, due to frequent distortions regarding the manual method, the (semi-)automated method seems to be more efficient and should be preferred.
评估使用两种不同软件(眼前节 OCT 内置软件和 MATLAB 自编软件)基于断层数据对储存于充满器官培养液 I(MI)或 II(MII)的塑料培养瓶中的供体角膜进行无菌角膜厚度测量的可靠性和效率。
使用眼前节 OCT 连续对 25 个储存于 MI 的供体角膜(50%)和 25 个储存于 MII 的供体角膜(50%)进行 5 次成像。使用眼前节 OCT 的手动测量工具(= CCTm)和允许(半自动)分析的 MATLAB 自编软件(= CCTa)分别测量中央角膜厚度(CCT)。我们使用 Cronbach's alpha(α)和 Wilcoxon 符号秩检验分析 CCTm 和 CCTa 的可靠性。
关于 CCTm,68 次测量(54.4%)在 MI 中存在 3D 体积成像扭曲,而 46 次测量(36.8%)在 MII 中存在扭曲,因此被排除。关于 CCTa,5 次测量(4%)在 MI 中,1 次测量(0.8%)在 MII 中无法分析。MI 中的平均 CCTm 为 1129±6.8μm,MII 中的平均 CCTm 为 820±5.1μm。MI 中的平均 CCTa 为 1149±2.7μm,MII 中的平均 CCTa 为 811±2.4μm。两种方法均具有较高的可靠性,CCTm 的 Cronbach's α 为 1.0(MI/MII),CCTa 的 Cronbach's α 为 0.99(MI)和 1.0(MII)。然而,在 MI 中,CCTm 的 5 次测量的平均标准差明显高于 CCTa(p=0.03),而在 MII 中则无差异(p=0.92)。
无菌供体断层摄影术可高度可靠地评估 CCT,两种方法均适用。然而,由于手动方法经常出现扭曲,因此(半自动)方法似乎更有效,应优先使用。