Suppr超能文献

Dymind D7-CRP 自动化分析仪的分析验证。

Analytical verification of the Dymind D7-CRP automated analyser.

机构信息

Department of Laboratory Diagnostics, Children's Hospital Zagreb, Zagreb, Croatia.

出版信息

Biochem Med (Zagreb). 2023 Jun 15;33(2):020703. doi: 10.11613/BM.2023.020703.

Abstract

INTRODUCTION

The aim of this study was to perform a verification of the Dymind D7-CRP automated analyser and compare it with established analysers.

MATERIALS AND METHODS

Analytical verification included estimation of repeatability, between run precision, within-laboratory precision, and bias in control samples with low, normal and high levels. The acceptance criteria for analytical verification were defined using the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) 2019 Biological Variation Database. Method comparison between the Dymind D7-CRP and the Sysmex XN1000 for haematological parameters and the Dymind D7-CRP and the Beckman Coulter AU680 for CRP values was performed on 40 patient samples.

RESULTS

Analytical verification criteria were adequately met with the exception of monocyte count for repeatability and within-laboratory precision (13.4% and 11.5%, respectively, acceptance criteria 10.1%) and measurement uncertainty (23.0, acceptance criteria 20.0%) at low level, eosinophil count for BIAS at the low level (37.7%, acceptance criteria 25.2%), basophil count (BAS) for BIAS at the high level (14.2%, acceptance criteria 10.9%), and mean platelet volume (MPV) for repeatability (4.2% and 6.8%), between run precision (2.2% and 4.7%), within-laboratory precision (4.0% and 7.3%) (acceptance criteria 1.7%), and measurement uncertainty (8.0 and 14.6%, acceptance criteria 3.4%) at both the low and high concentrations. Method comparison showed no clinically significant constant or proportional differences for all parameters except BAS and MPV.

CONCLUSION

The analytical verification of the Dymind D7-CRP showed adequate analytical characteristics. The Dymind D7-CRP can be used interchangeably with the Sysmex XN-1000 for all parameters tested, except BAS and MPV, and with the Beckman Coulter AU-680 for the determination of CRP.

摘要

简介

本研究的目的是对 Dymind D7-CRP 自动化分析仪进行验证,并将其与已建立的分析仪进行比较。

材料与方法

分析验证包括对重复性、批间精密度、实验室内部精密度以及低值、正常值和高值质控样本的偏倚进行评估。分析验证的接受标准使用 2019 年欧洲临床化学和实验室医学联合会(EFLM)生物学变异数据库进行定义。对 40 份患者样本进行了 Dymind D7-CRP 与 Sysmex XN1000 之间的血液学参数比较以及 Dymind D7-CRP 与 Beckman Coulter AU680 之间的 CRP 值比较。

结果

除了单核细胞计数的重复性和实验室内部精密度(分别为 13.4%和 11.5%,接受标准为 10.1%)和测量不确定度(23.0%,接受标准为 20.0%)在低值、嗜酸性粒细胞计数的低值偏倚(37.7%,接受标准为 25.2%)、嗜碱性粒细胞计数(BAS)的高值偏倚(14.2%,接受标准为 10.9%)以及血小板平均体积(MPV)的重复性(4.2%和 6.8%)、批间精密度(2.2%和 4.7%)、实验室内部精密度(4.0%和 7.3%)(接受标准为 1.7%)以及低值和高值时的测量不确定度(8.0%和 14.6%,接受标准为 3.4%)外,其他所有参数的分析验证标准均符合要求。方法比较显示,除 BAS 和 MPV 外,所有参数均无临床显著的恒定或比例差异。

结论

Dymind D7-CRP 的分析验证显示出足够的分析特性。Dymind D7-CRP 可与 Sysmex XN-1000 互换用于所有测试参数,除 BAS 和 MPV 外,还可与 Beckman Coulter AU-680 一起用于 CRP 的测定。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/d7f3/10231769/dbe8a41184cc/bm-33-2-020703-f1.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验