Diao Kai, Huang Xiaomin, Yao Mingyu, Li Jiawei, Fan Feifei, Pan Hongxian, Yu Jinjin, Yang Yizhou, Lu Weiwei, Lian Hengli, Wang Qinmei, Huang Jinhai, Chen Ruru
School of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical University, Wenzhou, China.
Eye Institute and Department of Ophthalmology, Eye & ENT Hospital, Fudan University, Shanghai, China.
Front Med (Lausanne). 2023 Jun 1;10:1161942. doi: 10.3389/fmed.2023.1161942. eCollection 2023.
This study aimed to test the inter-examiner and intra-examiner reliability of macular vascular density (VD) measurement of retinal and choriocapillaris plexuses in healthy children using optical coherence tomography angiography (OCTA).
Ninety-two school children were prospectively recruited. Macular OCTA images (6 × 6 mm) were obtained thrice by two examiners using the RTVue-XR Avanti OCT system. The coefficient of variation (COV), intraclass correlation coefficient (ICC), and Bland-Altman plots were used to evaluate the repeatability and reproducibility.
Ninety participants aged 6-15 years were enrolled; two participants were excluded because of low-quality images. In the retina, the reproducibility and repeatability of VD became poorer from superficial to deep retinal capillary plexus (superficial: COV = 4.61-11.11%; intermediate: COV = 7.73-14.15%; deep: COV = 14.60-32.28%). For both reproducibility and repeatability, the ICC ranged from moderate to high (superficial plexus: ICC = 0.570-0.976; intermediate plexus: ICC = 0.720-0.968; deep plexus: ICC = 0.628-0.954). In the choroid, the inter-examiner reproducibility and intra-examiner repeatability of the VD measurement of choriocapillaris were excellent in the macula, fovea, parafovea, and perifovea (COV = 1.00-6.10%; ICC = 0.856-0.950). The parameters of the foveal avascular zone (FAZ) also showed significant reproducibility and repeatability (COV = 0.01-0.21%; ICC = 0.743-0.994).
The VD measurements of the choriocapillaris and FAZ parameters using OCTA showed excellent inter-examiner and intra-examiner reliability in school children. The reproducibility and repeatability of the VD of three retinal capillary plexuses depended on the depth of the retinal capillary plexus.
本研究旨在通过光学相干断层扫描血管造影(OCTA)检测健康儿童视网膜和脉络膜毛细血管丛黄斑血管密度(VD)测量的检查者间和检查者内可靠性。
前瞻性招募92名学童。两名检查者使用RTVue-XR Avanti OCT系统对黄斑OCTA图像(6×6 mm)进行三次采集。采用变异系数(COV)、组内相关系数(ICC)和Bland-Altman图评估重复性和再现性。
纳入90名6至15岁的参与者;两名参与者因图像质量低而被排除。在视网膜中,从视网膜浅表层到深层毛细血管丛,VD的再现性和重复性逐渐变差(浅表层:COV = 4.61 - 11.11%;中间层:COV = 7.73 - 14.15%;深层:COV = 14.60 - 32.28%)。对于再现性和重复性,ICC范围从中等到高(浅表层毛细血管丛:ICC = 0.570 - 0.976;中间层毛细血管丛:ICC = 0.720 - 0.968;深层毛细血管丛:ICC = 0.628 - 0.954)。在脉络膜中,脉络膜毛细血管VD测量的检查者间再现性和检查者内重复性在黄斑、中央凹、旁中央凹和中央凹周围均极佳(COV = 1.00 - 6.10%;ICC = 0.856 - 0.950)。中央凹无血管区(FAZ)参数也显示出显著的再现性和重复性(COV = 0.01 - 0.21%;ICC = 0.743 - 0.994)。
使用OCTA测量脉络膜毛细血管和FAZ参数的VD在学童中显示出良好的检查者间和检查者内可靠性。三个视网膜毛细血管丛VD的再现性和重复性取决于视网膜毛细血管丛的深度。