Suppr超能文献

使用光学相干断层扫描血管造影术测量病理性眼睛中三个视网膜神经丛层血管密度的可重复性(OCTA三个神经丛层的血管密度可重复性)

Repeatability of Vascular Density Measurement of the Three Retinal Plexus Layers Using OCT Angiography in Pathologic Eyes (OCTA Vascular Density Repeatability of Three Plexus Layers).

作者信息

Mukkamala Lekha, Nguyen Michael, Chang Melinda, Park Susanna S

机构信息

Department of Ophthalmology & Vision Science, University of California Davis Eye Center, Sacramento, CA, USA.

Vision Center, Children's Hospital of Los Angeles, Los Angeles, CA, USA.

出版信息

Clin Ophthalmol. 2021 Jan 8;15:93-103. doi: 10.2147/OPTH.S284872. eCollection 2021.

Abstract

PURPOSE

Although commercial optical coherence tomography angiography (OCTA) machines quantitate retinal vascular density (VD) by dividing the vasculature into superficial and deep capillary plexus (SCP, DCP), histology reveals three distinct plexus layers. This study tested the hypothesis that the VD measurement of three distinct retinal plexus layers obtained using custom segmentation has high repeatability comparable to that of automatically segmented SCP and DCP layers.

MATERIALS AND METHODS

Forty-four participants (86 eyes) were enrolled - 54 eyes with retinal vasculopathy and 25 eyes with macular edema. Macular OCTA images (3x3 mm and 6x6 mm) were obtained twice within 30 minutes by the same personnel using the same instrument (AngioVue, Optovue, version 2018.0.0.18). The intraclass correlation coefficient (ICC) was calculated to access repeatability.

RESULTS

The repeatability of VD for SCP and DCP was good-to-moderate (ICC=0.65-0.85) and minimally affected by image quality, retinal vasculopathy, or macular edema. The repeatability of the VD of the custom-segmented intermediate and deep plexus layers (cICP and cDCP) was poor/moderate (ICC=0.40-0.74) but better in the subset without macular edema using 3x3 mm scans with good images quality (ICC=0.58-0.93). Repeatability of cICP and cDCP VD measurement for 6x6 mm scans was poor (ICC≤0.5) in eyes with retinal vasculopathy and/or macular edema.

CONCLUSION

Although repeatability of the VD measurement is high for the automatically segmented SCP and DCP, repeatability of VD is poor for the cICP and cDCP using larger scans in eyes with retinal vasculopathy and/or macular edema.

摘要

目的

尽管商用光学相干断层扫描血管造影(OCTA)机器通过将脉管系统分为浅层和深层毛细血管丛(SCP、DCP)来定量视网膜血管密度(VD),但组织学显示有三个不同的丛层。本研究检验了这样一个假设,即使用自定义分割获得的三个不同视网膜丛层的VD测量具有与自动分割的SCP和DCP层相当的高重复性。

材料与方法

招募了44名参与者(86只眼)——54只患有视网膜血管病变的眼睛和25只患有黄斑水肿的眼睛。黄斑OCTA图像(3×3毫米和6×6毫米)由同一人员在30分钟内使用同一台仪器(AngioVue,Optovue,版本2018.0.0.18)获取两次。计算组内相关系数(ICC)以评估重复性。

结果

SCP和DCP的VD重复性为中等至良好(ICC = 0.65 - 0.85),且受图像质量、视网膜血管病变或黄斑水肿的影响最小。自定义分割的中间和深层丛层(cICP和cDCP)的VD重复性为差/中等(ICC = 0.40 - 0.74),但在没有黄斑水肿的子集中,使用3×3毫米且图像质量良好的扫描时重复性较好(ICC = 0.58 - 0.93)。在患有视网膜血管病变和/或黄斑水肿的眼睛中,6×6毫米扫描的cICP和cDCP的VD测量重复性较差(ICC≤0.5)。

结论

尽管自动分割的SCP和DCP的VD测量重复性很高,但在患有视网膜血管病变和/或黄斑水肿的眼睛中,使用较大扫描时,cICP和cDCP的VD重复性较差。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/d352/7802913/66266af8238e/OPTH-15-93-g0001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验