Abstract

Review of the radiographs of 882 consecutive metal-to-polyethylene hip replacements after a mean follow-up of 6.8 years showed 72 cases (8%) of aseptic femoral loosening, all of which showed a lucent zone or black line between metal and cement. In 97% of the series this appeared within two years of the arthroplasty. Progress of loosening was unpredictable and only 24% had significant symptoms. Improved technique with methylmethacrylate cement in Stanmore prostheses had reduced loosening after two years from 7.9% to 0.4%. Methods of detection of early loosening are discussed.

摘要

对882例连续金属对聚乙烯髋关节置换术后平均随访6.8年的X线片进行回顾,发现72例(8%)出现无菌性股骨松动,所有病例在金属与骨水泥之间均显示透亮区或黑线。在该系列病例中,97%在关节置换术后两年内出现上述情况。松动进展无法预测,仅有24%出现明显症状。采用斯坦莫尔假体并改进甲基丙烯酸甲酯骨水泥技术后,两年后松动率从7.9%降至0.4%。文中讨论了早期松动的检测方法。