Effect of cardiac output on extravascular lung water measurements made with the thermoconductivity method.

作者信息

Noble W H, Kay J C

出版信息

Anesthesiology. 1986 Aug;65(2):206-9. doi: 10.1097/00000542-198608000-00014.

Abstract

The authors tested the effects of injectate temperature and of changing cardiac output on the measurement of lung water by the thermoconductivity technique (ETVL). Cardiac output in dogs was increased and decreased with isoproterenol and halothane, respectively. Post mortem extravascular lung water (pulmonary extravascular tissue weight [PETW] was determined using a weighing technique. Cardiac output varied between 0.7 and 8.8 l/min and did not influence the ETVL measurement. The authors conclude ETVL measured by the thermoconductivity technique is not influenced by large changes in cardiac output.

摘要

作者通过热导率技术(ETVL)测试了注入液温度及心输出量变化对肺水测量的影响。分别使用异丙肾上腺素和氟烷使犬的心输出量增加和减少。采用称重技术测定死后血管外肺水(肺血管外组织重量[PETW])。心输出量在0.7至8.8升/分钟之间变化,且不影响ETVL测量。作者得出结论,热导率技术测量的ETVL不受心输出量大幅变化的影响。