J Refract Surg. 2023 Jul;39(7):474-480. doi: 10.3928/1081597X-20230606-01. Epub 2023 Jul 1.
To assess repeatability and agreement of corneal epithelial thickness mapping in eyes with keratoconus using three optical coherence tomography (OCT) devices featuring different technologies: spectral-domain (SD) OCT combined with Placido disk corneal topography (MS-39), swept-source OCT (Anterion), and SD-OCT (Avanti).
Three consecutive measurements were acquired with the three devices in 60 eyes with keratoconus. The mean epithelial thickness was calculated in the central 2-mm zone and in 2- to 5-mm and 5- to 7-mm diameter rings. The repeatability was calculated using pooled within-subject standard deviation (S). The agreement was assessed by paired tests and Bland-Altman plots.
The repeatability (Sw) of the epithelial thickness for the central 2-mm zone was 0.91, 0.71, and 0.93 μm for the MS-39, Anterion, and Avanti, respectively. All thicknesses with the MS-39 were greater than those of the Anterion and Avanti, with mean differences of 4.11 ± 1.34 μm ( < .001) and 0.52 ± 1.30 μm ( = .003), respectively. The 95% limits of agreement were 1.484 to 6.736 μm for the MS-39 and Anterion, -3.068 to 2.028 μm for the Avanti and MS-39, and 1.258 to 5.922 μm for for the Avanti and Anterion.
Epithelial thickness mapping results were most repeatable with the Anterion, followed by the MS-39 and Avanti. The MS-39 gave the thickest values, followed by the Avanti and Anterion. The differences were significant, making the devices not interchangeable for epithelial thickness mapping in eyes with keratoconus. .
使用三种具有不同技术的光学相干断层扫描(OCT)设备评估圆锥角膜患者角膜上皮厚度测绘的可重复性和一致性,这些设备包括:谱域(SD)OCT 结合角膜地形图(MS-39)、扫频源 OCT(Anterion)和 SD-OCT(Avanti)。
对 60 只圆锥角膜眼的三种设备进行了三次连续测量。在中央 2mm 区域和 2 至 5mm 和 5 至 7mm 直径环中计算平均上皮厚度。使用总体个体内标准差(S)计算重复性。通过配对检验和 Bland-Altman 图评估一致性。
中央 2mm 区域上皮厚度的重复性(Sw)分别为 MS-39、Anterion 和 Avanti 的 0.91、0.71 和 0.93μm。MS-39 的所有厚度均大于 Anterion 和 Avanti,平均差异分别为 4.11±1.34μm(<0.001)和 0.52±1.30μm(=0.003)。MS-39 和 Anterion 的 95%一致性区间分别为 1.484 至 6.736μm,Avanti 和 MS-39 为-3.068 至 2.028μm,Avanti 和 Anterion 为 1.258 至 5.922μm。
上皮厚度测绘结果以 Anterion 最为可靠,其次是 MS-39 和 Avanti。MS-39 给出的厚度最大,其次是 Avanti 和 Anterion。这些差异具有统计学意义,因此这些设备不能在圆锥角膜患者的上皮厚度测绘中互换使用。