Feng Yue, Reinstein Dan Z, Nitter Tore, Archer Timothy J, McAlinden Colm, Chen Xiangjun, Bertelsen Geir, Utheim Tor Paaske, Stojanovic Aleksandar
J Refract Surg. 2022 Jun;38(6):356-363. doi: 10.3928/1081597X-20220414-01. Epub 2022 Jun 1.
To assess the repeatability of corneal epithelial thickness mapping in virgin, post-laser refractive surgery (PLRS), and keratoconic eyes using a novel swept-source optical coherence tomographer (SS-OCT), and to determine the agreement of the measurements with a validated spectral-domain (SD) OCT.
Analysis of 90 virgin, 46 PLRS, and 122 keratoconic eyes was performed. Three consecutive measurements of each eye were acquired with the Anterion SS-OCT and Avanti SD-OCT devices, and averages of the epithelial thickness mapping were calculated in the central 2-mm zone and in the 2- to 5-mm and 5- to 7-mm diameter rings. The repeatability was analyzed using pooled within-subject standard deviation (S). The agreement was assessed by Bland-Altman analysis and paired tests.
The repeatability ranges of the Anterion and Avanti epithelial thickness mapping measurements were S: 0.60 to 1.36 µm and S: 0.75 to 1.96 µm, respectively. The 95% limits of agreement of the Anterion and Avanti were 0.826 to 8.297. All values of the thickness measurements with the Anterion were lower than those of the Avanti, with the mean differences being 4.06 ± 1.81, 3.26 ± 2.52, and 3.68 ± 2.51 µm in virgin, PLRS, and keratoconic eyes, respectively ( < .001 for all).
The repeatability of the Anterion's epithelial thickness mapping was higher than that of the Avanti. In terms of the agreement between the Anterion and Avanti, the epithelium measured by the Anterion was always thinner than that of the Avanti, making their interchangeable use unsuitable without corrections. .
使用新型扫频光学相干断层扫描仪(SS-OCT)评估初次检查、激光屈光性手术后(PLRS)和圆锥角膜患者角膜上皮厚度测量的可重复性,并确定测量值与经过验证的谱域(SD)OCT的一致性。
对90只初次检查眼、46只PLRS眼和122只圆锥角膜眼进行分析。使用Anterion SS-OCT和Avanti SD-OCT设备对每只眼睛进行连续三次测量,并计算中央2mm区域以及直径2至5mm和5至7mm环形区域上皮厚度测量的平均值。使用合并的受试者内标准差(S)分析可重复性。通过Bland-Altman分析和配对检验评估一致性。
Anterion和Avanti上皮厚度测量的可重复性范围分别为S:0.60至1.36μm和S:0.75至1.96μm。Anterion和Avanti的95%一致性界限为0.826至8.297。Anterion测量的所有厚度值均低于Avanti,初次检查眼、PLRS眼和圆锥角膜眼中的平均差异分别为-4.06±1.81、-3.26±2.52和-3.68±2.51μm(所有均P<0.001)。
Anterion上皮厚度测量的可重复性高于Avanti。就Anterion和Avanti之间的一致性而言,Anterion测量的上皮始终比Avanti测量的薄,因此未经校正不宜互换使用。