Repeatability of Automatic Measurements by an Anterior Segment Swept-Source OCT Biometer in Patients With Keratoconus.

出版信息

J Refract Surg. 2024 Jul;40(7):e445-e452. doi: 10.3928/1081597X-20240514-03. Epub 2024 Jul 1.

Abstract

PURPOSE

To evaluate the repeatability of automatic measurements of a new anterior segment optical coherence tomographer (ANTERION; Heidelberg Engineering) and their agreement with an anterior segment optical coherence tomography device combined with Placido disc corneal topography (MS-39; CSO) in patients affected by keratoconus.

METHODS

Fifty-four consecutive patients were included. Three measurements were performed with the ANTERION and one with the MS-39. Repeatability was assessed by means of within-subject standard deviation, coefficient of variation (CoV), and intraclass correlation coefficient (ICC). Agreement was investigated with the 95% limits of agreement. The paired -test and Wilcoxon matched-pairs test were performed to compare the measurements of the different devices.

RESULTS

Repeatability of ANTERION measurements was high, with an ICC greater than 0.98 for all parameters. Many parameters revealed a CoV of less than 1% and a CoV within 5% was obtained for astigmatism measurements. The ANTERION measured a significantly higher corneal power and the MS-39 more negative posterior keratometric values. These differences were mirrored by a moderate agreement for mean simulated keratometry and poor agreement for total corneal power and posterior keratometry.

CONCLUSIONS

The ANTERION revealed high repeatability of automatic measurements and good agreement with the MS-39 for many parameters in patients affected by keratoconus, but for most parameters the two instruments cannot be considered interchangeable. .

摘要

目的

评估新型眼前节光学相干断层扫描仪(ANTERION;Heidelberg Engineering)自动测量的重复性及其与结合角膜 Placido 盘地形图的眼前节光学相干断层扫描仪(MS-39;CSO)在圆锥角膜患者中的一致性。

方法

纳入 54 例连续患者。使用 ANTERION 进行 3 次测量,使用 MS-39 进行 1 次测量。通过个体内标准差、变异系数(CoV)和组内相关系数(ICC)评估重复性。使用 95%一致性界限研究一致性。采用配对 t 检验和 Wilcoxon 配对检验比较两种设备的测量值。

结果

ANTERION 测量的重复性较高,所有参数的 ICC 均大于 0.98。许多参数的 CoV 小于 1%,散光测量的 CoV 在 5%以内。ANTERION 测量的角膜曲率较高,而 MS-39 测量的后角膜曲率值较负。这些差异反映在平均模拟角膜曲率的中等一致性和总角膜曲率和后角膜曲率的较差一致性上。

结论

在圆锥角膜患者中,ANTERION 对自动测量具有较高的重复性,与 MS-39 具有良好的一致性,但对于大多数参数,两种仪器不能互换。