The purpose of this study was to develop a concept for predicting the effects of both discrete intraocular lens (IOL) power steps (PS) and power labelling tolerances (LT) on the uncertainty of the refractive outcome (REFU).
Retrospective non-randomised cross-sectional Monte Carlo simulation study.
We evaluated a dataset containing 16 669 IOLMaster 700 preoperative biometric measurements. The PS and the delivery range of two modern IOLs (Bausch and Lomb enVista and Alcon SA60AT) were considered for this Monte Carlo simulation. The uncertainties from PS or LT were assumed to be normally distributed according to ±½ the IOL PS or the ISO 11979 LT. REFU was recorded and analysed for all simulations.
With both lenses the REFU from discrete PS ranged from 0.11 to 0.12 dpt. Due to the larger PS for low/high power lenses with the enVista/SA60AT, REFU is more dominant in initially myopic/hyperopic eyes. REFU from LT ranged from 0.18 to 0.19 dpt for both lenses. Since LT increases stepwise with IOL power, REFU is more prevalent in initially hyperopic eyes requiring high IOL power values, and for lenses with a wide delivery range towards higher powers.
Since surgeons and patients are typically aware of the effect of discrete PS on REFU, these might be tolerated in cataract surgery. However, REFU resulting from LT is inevitable while the true measured IOL power is not reported on the package, leading to background noise in postoperative achieved refraction.
本研究旨在提出一种概念,以预测离散人工晶状体(IOL)功率步长(PS)和功率标签公差(LT)对屈光结果不确定性(REFU)的影响。
回顾性非随机横断面蒙特卡罗模拟研究。
我们评估了包含 16669 例 IOLMaster 700 术前生物测量数据的数据集。对于这项蒙特卡罗模拟,考虑了两种现代 IOL(Bausch and Lomb enVista 和 Alcon SA60AT)的 PS 和输送范围。假设 PS 或 LT 的不确定性符合 PS 的 ±½ IOL 或 ISO 11979 LT。记录和分析了所有模拟的 REFU。
对于两种镜片,离散 PS 的 REFU 范围为 0.11 至 0.12 dpt。由于 enVista/SA60AT 具有较低/较高功率镜片的较大 PS,REFU 在最初近视/远视眼中更为明显。两种镜片的 LT 的 REFU 范围为 0.18 至 0.19 dpt。由于 LT 随着 IOL 功率的逐步增加,REFU 在最初需要高 IOL 功率值的远视眼中更为常见,并且对于具有宽输送范围的镜片,其功率值较高。
由于外科医生和患者通常都意识到离散 PS 对 REFU 的影响,因此在白内障手术中可能会容忍这些影响。然而,由于在包装上未报告真实测量的 IOL 功率,因此 LT 导致的 REFU 是不可避免的,这会导致术后实际屈光度的背景噪声。