Evaluation of the accuracy of intraoral scanners for complete-arch scanning: A systematic review and network meta-analysis.

机构信息

Centre for Translational Medicine, Semmelweis University, Budapest, Hungary; Department of Prosthodontics, Semmelweis University, Budapest, Hungary.

Centre for Translational Medicine, Semmelweis University, Budapest, Hungary; Department of Periodontology, Semmelweis University, Budapest, Hungary.

出版信息

J Dent. 2023 Oct;137:104636. doi: 10.1016/j.jdent.2023.104636. Epub 2023 Jul 27.

Abstract

OBJECTIVES

This network meta-analysis (NMA) aimed to compare the complete-arch scanning accuracy of different intraoral scanners (IOSs) to that of reference standard tessellation language (STL) files.

DATA

Studies comparing the trueness and precision of IOS STL files with those of reference STL scans for different arch types (dentate, edentulous, completely edentulous with implants, and partially edentulous with implants) were included in this study.

SOURCES

An electronic search of five databases restricted to the English Language was conducted in October 2021.

STUDY SELECTION

A total of 3,815 studies were identified, of which 114 were eligible for inclusion. After study selection and data extraction, pair-wise comparison and NMA were performed to define the accuracy of scanning for four arch subgroups using four outcomes (trueness and precision expressed as mean absolute deviation and root mean square values). Cochrane guidelines and the QUADAS-2 tool were used to assess the risk of bias. GRADE was used for certainty assessment.

RESULTS

Fifty-three articles were included in this NMA. Altogether, 26 IOSs were compared directly and indirectly in 10 network systems. The accuracy of IOSs scans were not significantly different from the reference scans for dentate arches (three IOSs), edentulous arches (three IOSs), and completely edentulous arches with implants (one IOS). The accuracy of the IOSs was significantly different from the reference scans for partially edentulous arches with implants. Significant accuracy differences were found between the IOSs, regardless of clinical scenarios.

CONCLUSIONS

The accuracy of complete-arch scanning by IOSs differs based on clinical scenarios.

CLINICAL SIGNIFICANCE

Different IOSs should be used according to the complete arch type.

摘要

目的

本网络荟萃分析(NMA)旨在比较不同口内扫描仪(IOS)的全牙弓扫描准确性与参考标准体素语言(STL)文件的准确性。

资料来源

本研究纳入了比较不同牙弓类型(有牙、无牙、全口种植义齿无牙和部分种植义齿无牙)的 IOS STL 文件与参考 STL 扫描的准确性的研究。

检索

2021 年 10 月,我们对五个数据库进行了电子检索,限制使用英语。

研究选择

共确定了 3815 项研究,其中有 114 项符合纳入标准。在进行研究选择和数据提取后,我们进行了两两比较和 NMA,以使用四个结果(以均方根值和平均绝对偏差表示的准确性和精密度)定义四个牙弓亚组的扫描准确性。我们使用 Cochrane 指南和 QUADAS-2 工具评估偏倚风险。我们使用 GRADE 评估确定性。

结果

这项 NMA 纳入了 53 篇文章。总共在 10 个网络系统中直接和间接比较了 26 种 IOS。对于有牙弓(三种 IOS)、无牙弓(三种 IOS)和全口种植义齿无牙弓(一种 IOS),IOS 扫描的准确性与参考扫描无显著差异。对于部分种植义齿无牙弓,IOS 的准确性与参考扫描有显著差异。无论临床情况如何,都发现 IOS 之间存在显著的准确性差异。

结论

IOS 全牙弓扫描的准确性取决于临床情况。

临床意义

应根据全牙弓类型选择不同的 IOS。