To test and initially describe a new handheld wireless ultrasound technique (TE Air) for clinical use.
In this pilot study, the new ultrasound device TE Air from Mindray was used to examine the hepatic and renal vessels of healthy volunteers for first impressions. The probe has a sector transducer with a frequency range of 1.8-4.5 MHz. The B-mode and color-coded doppler sonography (CCDS) scanning methods were used. A high-end device from the same company (Resona 9, Mindray) was used as a reference. The results were evaluated using an image rating scale ranging from 0 to 5, with 0 indicating not assessable and 5 indicating without limitations.
Altogether, 61 participants (n = 34 female [55.7%], n = 27 male [44.3%]), age range 18-83 years, mean age 37.9±16.5 years) could be adequately studied using TE AIR and the high-end device. With one exception, the image quality score for TE Air never fell below 3 and had a mean/median scored of 4.97/5.00 for the B-mode, 4.92/5.00 for the color flow (CF) mode, and 4.89/5.00 for the pulse wave (PW) mode of the hepatic vein, 4.90/5.00 for the portal vein, 4.11/4.00 for the hepatic artery, and 4.57/5.00 for the renal segmental artery. A significant difference in the assessment of flow measurement of the hepatic artery and renal segmental arteries was found between TE AIR and the high-end device.
TE Air represents a new dimension in point-of-care ultrasound via wireless handheld devices. Especially, its flow measurement ability offers a relevant advantage over other available handheld models. TE Air provides a formally sufficient image quality in terms of diagnostic significance.
测试并初步描述一种新的手持式无线超声技术(TE Air)在临床中的应用。
在这项初步研究中,使用迈瑞公司的新型超声设备 TE Air 对健康志愿者的肝、肾血管进行初步检查。该探头采用扇型换能器,频率范围为 1.8-4.5MHz。采用 B 型模式和彩色编码多普勒超声(CCDS)扫描方法。使用同一家公司(迈瑞 Resona 9)的高端设备作为参考。使用 0 到 5 的图像评分量表对结果进行评估,0 表示无法评估,5 表示无限制。
共有 61 名参与者(n=34 名女性[55.7%],n=27 名男性[44.3%]),年龄 18-83 岁,平均年龄 37.9±16.5 岁)可以使用 TE Air 和高端设备进行充分研究。除了一个例外,TE Air 的图像质量评分从未低于 3,其 B 模式的平均/中位数评分为 4.97/5.00,彩色血流(CF)模式为 4.92/5.00,脉冲波(PW)模式为 4.89/5.00,肝静脉的门静脉为 4.90/5.00,肝动脉为 4.11/4.00,肾段动脉为 4.57/5.00。在肝动脉和肾段动脉的血流测量评估中,TE Air 和高端设备之间存在显著差异。
TE Air 通过无线手持式设备为即时护理超声技术提供了一个新的维度。特别是,其流量测量能力相对于其他可用的手持式模型具有重要优势。TE Air 在诊断意义方面提供了充分的图像质量。