Medical Clinic and Polyclinic III, University Hospital Regensburg, Regensburg, Germany.
Department of Medical Statistics, University Medical Center Göttingen, Göttingen, Germany.
Clin Hemorheol Microcirc. 2024;88(s1):S57-S68. doi: 10.3233/CH-248105.
To assess diagnostic capabilities of a modern handheld ultrasound device (HUD) for portal vein (PV) evaluation in pediatric patients and to verify if age, body mass index (BMI) and object depth correlate to imagequality.
45 patients (3.9 months- 17.9 years; 10.3 years±5.4) were examined using a HUD and cart-based high-end ultrasound system (HEUS). The intra-and extrahepatic PV was scanned using B-Mode and Color-coded Doppler sonography (CCDS). A five-point Likert scale was applied to evaluate image quality. Results were interpreted by two readers in consensus. Scores were compared between the scanners and correlated to age, BMI and skin-to-portal-vein-distance (SPVD). The influence on image quality and the difference between the two devices were analyzed using ordinal and Bayesian logistic regression models.
≥4 points (mild or no limitations) were achieved by the HUD in 60% for B-Mode and 56% for CCDS. There was a statistically significant negative correlation between the image quality achieved by the HUD and the SPVD for B-Mode and CCDS. The effect of BMI and SPVD on having≥4 points differed significantly between the HUD and HEUS for B-Mode and for CCDS.
Sufficient to excellent image quality for portal vein assessment in pediatric patients was achieved by the HUD except for the most deep-lying structures. Compared to HEUS, diagnostic performance of the HUD is lower, based on a negative correlation with object depth.
评估一种现代手持式超声设备(HUD)在儿科患者门静脉(PV)评估中的诊断能力,并验证年龄、体重指数(BMI)和目标深度与图像质量是否相关。
使用 HUD 和基于推车的高端超声系统(HEUS)对 45 名患者(3.9 个月至 17.9 岁;10.3 岁±5.4 岁)进行检查。使用 B 模式和彩色多普勒超声(CCDS)对肝内外门静脉进行扫描。采用 5 分李克特量表评估图像质量。由两名读者进行共识解读。对扫描仪之间的评分进行比较,并与年龄、BMI 和皮肤至门静脉距离(SPVD)相关联。使用有序和贝叶斯逻辑回归模型分析对图像质量的影响和两种设备之间的差异。
HUD 在 B 模式下获得≥4 分(轻度或无限制)的比例为 60%,在 CCDS 下为 56%。HUD 获得的图像质量与 B 模式和 CCDS 的 SPVD 呈统计学显著负相关。BMI 和 SPVD 对获得≥4 分的影响在 HUD 和 HEUS 之间在 B 模式和 CCDS 方面存在显著差异。
HUD 除了最深处的结构外,在儿科患者门静脉评估中可以获得足够好的图像质量。与 HEUS 相比,HUD 的诊断性能较低,这与目标深度呈负相关。