A full-thickness macular hole ("FTMH") is a foveal lesion caused by a defect in the full thickness of the neurosensory retina. Its diagnosis and the indication for surgical treatment take into account the measurement of the hole according to the tool provided by the OCT. This measurement can be performed by several ophthalmologists during the follow-up of a patient. The aim of this study is to find out whether there is intra-individual and inter-individual variability in these measurements.
Retrospective review of OCT b-scan images with a diagnosis of FTMH. Measurements of the minimum diameter of the FTMH were performed using the hand-held tool available on the DRI-Triton (Topcon, Japan) at 1:1 and 1:2 scales, on different days, by 2 retina specialists and 2 residents. These measurements were compared to assess inter-observer and intra-observer correspondence.
Thirty-four images were analysed. For intra-observer variability, a correlation index higher than 0.98 was obtained in all cases. For inter-observer variability, the intra-class correlation coefficient was 0.94 (95% CI: 0.91-0.97) for the 1:1 scale, and 0.94 (95% CI: 0.91-0.97) for the 1:2 scale.
OCT-measured AMEC size values are reproducible between ophthalmic specialists and residents and are independent of the imaging scale at which the measurement is made.
全层黄斑裂孔(“FTMH”)是一种由神经感觉视网膜全层缺陷引起的黄斑病变。其诊断和手术治疗指征考虑了根据 OCT 提供的工具对孔进行测量。在患者的随访过程中,可能会有多位眼科医生进行这种测量。本研究的目的是确定这些测量值是否存在个体内和个体间的差异。
回顾性分析诊断为 FTMH 的 OCT B 扫描图像。使用 DRI-Triton(Topcon,日本)上的手持工具,在 1:1 和 1:2 比例下,由 2 名视网膜专家和 2 名住院医师在不同日期对 FTMH 的最小直径进行测量。比较这些测量值以评估观察者间和观察者内的一致性。
分析了 34 张图像。对于观察者内的可变性,所有情况下的相关指数均高于 0.98。对于观察者间的可变性,在 1:1 比例下,组内相关系数为 0.94(95%置信区间:0.91-0.97),在 1:2 比例下为 0.94(95%置信区间:0.91-0.97)。
OCT 测量的 AMEC 大小值在眼科专家和住院医师之间具有可重复性,并且与进行测量的成像比例无关。