四维血流磁共振成像的血流评估软件:可靠性和验证研究。

Flow evaluation software for four-dimensional flow MRI: a reliability and validation study.

机构信息

Paediatric Cardiology, Pediatric Heart Center, Department of Surgery, University Children's Hospital Zürich, Steinwiesstrasse 75, 8032, Zurich, Switzerland.

Children's Research Center, University Children's Hospital Zürich, Zurich, Switzerland.

出版信息

Radiol Med. 2023 Oct;128(10):1225-1235. doi: 10.1007/s11547-023-01697-4. Epub 2023 Aug 24.

Abstract

PURPOSE

Four-dimensional time-resolved phase-contrast cardiovascular magnetic resonance imaging (4D flow MRI) enables blood flow quantification in multiple vessels, which is crucial for patients with congenital heart disease (CHD). We investigated net flow volumes in the ascending aorta and pulmonary arteries by four different postprocessing software packages for 4D flow MRI in comparison with 2D cine phase-contrast measurements (2D PC).

MATERIAL AND METHODS

4D flow and 2D PC datasets of 47 patients with biventricular CHD (median age 16, range 0.6-52 years) were acquired at 1.5 T. Net flow volumes in the ascending aorta, the main, right, and left pulmonary arteries were measured using four different postprocessing software applications and compared to offset-corrected 2D PC data. Reliability of 4D flow postprocessing software was assessed by Bland-Altman analysis and intraclass correlation coefficient (ICC). Linear regression of internal flow controls was calculated. Interobserver reproducibility was evaluated in 25 patients.

RESULTS

Correlation and agreement of flow volumes were very good for all software compared to 2D PC (ICC ≥ 0.94; bias ≤ 5%). Internal controls were excellent for 2D PC (r ≥ 0.95, p < 0.001) and 4D flow (r ≥ 0.94, p < 0.001) without significant difference of correlation coefficients between methods. Interobserver reliability was good for all vendors (ICC ≥ 0.94, agreement bias < 8%).

CONCLUSION

Haemodynamic information from 4D flow in the large thoracic arteries assessed by four commercially available postprocessing applications matches routinely performed 2D PC values. Therefore, we consider 4D flow MRI-derived data ready for clinical use in patients with CHD.

摘要

目的

四维时相对比心血管磁共振成像(4D flow MRI)可对多支血管的血流进行定量分析,这对先天性心脏病(CHD)患者至关重要。我们通过四种不同的 4D 流后处理软件包对 47 例双心室 CHD 患者的升主动脉和肺动脉的净血流量进行了研究,并与二维电影相位对比测量(2D PC)进行了比较。

材料与方法

在 1.5T 上采集了 47 例双心室 CHD 患者的 4D flow 和 2D PC 数据集(中位数年龄为 16 岁,范围为 0.6-52 岁)。使用四种不同的后处理软件应用程序测量升主动脉、主肺动脉、右肺动脉和左肺动脉的净血流量,并与偏移校正的 2D PC 数据进行比较。通过 Bland-Altman 分析和组内相关系数(ICC)评估 4D 流后处理软件的可靠性。计算内部流量控制的线性回归。在 25 例患者中评估了观察者间的可重复性。

结果

与 2D PC 相比,所有软件的流量相关性和一致性均非常好(ICC≥0.94;偏差≤5%)。2D PC(r≥0.95,p<0.001)和 4D flow(r≥0.94,p<0.001)的内部对照均非常好,两种方法之间的相关系数无显著差异。所有供应商的观察者间可靠性均良好(ICC≥0.94,一致性偏差<8%)。

结论

四种商业上可用的后处理应用程序评估的大型胸主动脉 4D 流的血液动力学信息与常规 2D PC 值相匹配。因此,我们认为 4D flow MRI 衍生的数据可以在 CHD 患者的临床应用中使用。