Suppr超能文献

使用2×2并行成像加速技术,在脉动流模体以及有和没有呼吸门控的活体中,对两家供应商的心血管4D流磁共振成像进行验证和可重复性研究。

Validation and reproducibility of cardiovascular 4D-flow MRI from two vendors using 2 × 2 parallel imaging acceleration in pulsatile flow phantom and in vivo with and without respiratory gating.

作者信息

Bock Jelena, Töger Johannes, Bidhult Sebastian, Markenroth Bloch Karin, Arvidsson Per, Kanski Mikael, Arheden Håkan, Testud Frederik, Greiser Andreas, Heiberg Einar, Carlsson Marcus

机构信息

1 Department of Clinical Sciences, Lund University, Clinical Physiology, Skåne University Hospital, Lund, Sweden.

2 Department of Diagnostic Radiology, Lund University, Skåne University Hospital, Lund, Sweden.

出版信息

Acta Radiol. 2019 Mar;60(3):327-337. doi: 10.1177/0284185118784981. Epub 2018 Jun 26.

Abstract

BACKGROUND

4D-flow magnetic resonance imaging (MRI) is increasingly used.

PURPOSE

To validate 4D-flow sequences in phantom and in vivo, comparing volume flow and kinetic energy (KE) head-to-head, with and without respiratory gating.

MATERIAL AND METHODS

Achieva dStream (Philips Healthcare) and MAGNETOM Aera (Siemens Healthcare) 1.5-T scanners were used. Phantom validation measured pulsatile, three-dimensional flow with 4D-flow MRI and laser particle imaging velocimetry (PIV) as reference standard. Ten healthy participants underwent three cardiac MRI examinations each, consisting of cine-imaging, 2D-flow (aorta, pulmonary artery), and 2 × 2 accelerated 4D-flow with (Resp+) and without (Resp-) respiratory gating. Examinations were acquired consecutively on both scanners and one examination repeated within two weeks. Volume flow in the great vessels was compared between 2D- and 4D-flow. KE were calculated for all time phases and voxels in the left ventricle.

RESULTS

Phantom results showed high accuracy and precision for both scanners. In vivo, higher accuracy and precision ( P < 0.001) was found for volume flow for the Aera prototype with Resp+ (-3.7 ± 10.4 mL, r = 0.89) compared to the Achieva product sequence (-17.8 ± 18.6 mL, r = 0.56). 4D-flow Resp- on Aera had somewhat larger bias (-9.3 ± 9.6 mL, r = 0.90) compared to Resp+ ( P = 0.005). KE measurements showed larger differences between scanners on the same day compared to the same scanner at different days.

CONCLUSION

Sequence-specific in vivo validation of 4D-flow is needed before clinical use. 4D-flow with the Aera prototype sequence with a clinically acceptable acquisition time (<10 min) showed acceptable bias in healthy controls to be considered for clinical use. Intra-individual KE comparisons should use the same sequence.

摘要

背景

四维血流磁共振成像(MRI)的应用日益广泛。

目的

在体模和体内验证四维血流序列,在有和没有呼吸门控的情况下,直接比较容积流量和动能(KE)。

材料与方法

使用了飞利浦医疗的Achieva dStream和西门子医疗的MAGNETOM Aera 1.5-T扫描仪。体模验证采用四维血流MRI测量脉动三维血流,并以激光粒子成像测速技术(PIV)作为参考标准。10名健康受试者每人接受了三次心脏MRI检查,包括电影成像、二维血流(主动脉、肺动脉)以及有(Resp+)和无(Resp-)呼吸门控的2×2加速四维血流检查。在两台扫描仪上连续进行检查,且在两周内重复一次检查。比较二维血流和四维血流中大血管的容积流量。计算左心室所有时间相位和体素的动能。

结果

体模结果显示两台扫描仪均具有较高的准确性和精密度。在体内,与Achieva产品序列(-17.8±18.6 mL,r = 0.56)相比,配备Resp+的Aera原型机在容积流量方面具有更高的准确性和精密度(P < 0.001,-3.7±10.4 mL,r = 0.89)。与Resp+相比,Aera上的四维血流Resp-偏差稍大(-9.3±9.6 mL,r = 0.90,P = 0.005)。与同一台扫描仪在不同日期相比,同一天不同扫描仪之间的动能测量差异更大。

结论

在临床应用前,需要对四维血流进行特定序列的体内验证。配备临床可接受采集时间(<10分钟)的Aera原型序列的四维血流在健康对照中显示出可接受的偏差,可考虑用于临床。个体内的动能比较应使用相同序列。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/df86/6402051/3ca40b12c4ab/10.1177_0284185118784981-fig1.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验