Dow C S, France A D, Khan M S, Johnson T
J Clin Pathol. 1979 Apr;32(4):386-90. doi: 10.1136/jcp.32.4.386.
The accuracy and practicality of particle size distribution analysis for rapid screening of urine specimens are assessed. Six hundred urines were subjected simultaneously to routine bacteriological examinations and particle size distribution analysis using a Coulter Counter (ZBI) linked to a C1000 Channelyzer. There was complete agreement in the results of 593 (98.8%) specimens. Characteristic profiles of various bacterial species in infected specimens were consistently obtained. This system can easily be linked to any existing computer reporting in a district hospital laboratory, and the results of negative specimens (70--80%) can be obtained within 5--10 minutes.
评估了用于尿液标本快速筛查的粒度分布分析的准确性和实用性。600份尿液同时进行常规细菌学检查,并使用与C1000通道分析仪相连的库尔特计数器(ZBI)进行粒度分布分析。593份(98.8%)标本的结果完全一致。在感染标本中始终获得了各种细菌种类的特征图谱。该系统可以很容易地与地区医院实验室现有的任何计算机报告系统相连,阴性标本(70%-80%)的结果可在5-10分钟内获得。