研究单源多层计算机轴向断层扫描的层厚效应对噪声和诊断信息量的影响。

Investigating the slice thickness effect on noise and diagnostic content of single-source multi-slice computerized axial tomography.

机构信息

Department of Pharmacology and Medical Physics and Clinical Biochemistry, College of Medicine, Hawler Medical University, Erbil, Iraq.

Department of Radiology, Erbil Technical Medical Institute, Erbil Polytechnic University, Erbil, Iraq.

出版信息

J Med Life. 2023 Jun;16(6):862-867. doi: 10.25122/jml-2022-0188.

Abstract

High-quality and detailed CT scan images are crucial for accurate diagnosis. Factors such as image noise and slice thickness affect image quality. This study aimed to determine the optimal slice thickness that minimized image noise while maintaining sufficient diagnostic information using the single-source computed tomography head protocol. Single-source CT images were examined using the Linux Operating system Ge Revolution 64-slice CT scanner, and a combination of statical analysis and DICOM CT image analysis was employed. The single-source energy head CT protocol was used to investigate the effect of slice thickness on noise and visibility in images. Different values of slice thickness 0.625, 1.25, 2.5, 3.75, 5, 7.5, and 10 were prepared, and then quantitative analysis was performed. Thinner slice thickness decreased image noise, increased visibility, and improved detection. Therefore, the balance between changing the thickness of the slice with the diagnostic content and image noise must be considered. Maximum slice thickness enhances CT image detail and structure despite more noise. Based on the results, a slice thickness of 1.25mm was identified as the optimal choice for reducing image noise and achieving better and more accurate detection using the single-source computed tomography head protocol. The study revealed that image noise tends to increase with greater slice thickness according to the Linux operating system. These findings can serve as a valuable guide for quality control methods in CT centers, emphasizing the need to determine the appropriate slice thickness to ensure an accurate diagnosis.

摘要

高质量且详细的 CT 扫描图像对于准确诊断至关重要。图像噪声和切片厚度等因素会影响图像质量。本研究旨在使用单源计算机断层头部协议确定最小化图像噪声同时保持足够诊断信息的最佳切片厚度。使用 Linux 操作系统的 Ge Revolution 64 层 CT 扫描仪检查单源 CT 图像,并结合静态分析和 DICOM CT 图像分析。使用单源能量头部 CT 协议研究切片厚度对图像噪声和可见度的影响。准备了不同的切片厚度值 0.625、1.25、2.5、3.75、5、7.5 和 10,并进行了定量分析。较薄的切片厚度降低了图像噪声,提高了可见度,改善了检测效果。因此,必须在改变切片厚度与诊断内容和图像噪声之间取得平衡。尽管会增加噪声,但最大切片厚度可增强 CT 图像的细节和结构。基于这些结果,确定 1.25mm 的切片厚度是使用单源计算机断层头部协议降低图像噪声并实现更好、更准确检测的最佳选择。研究表明,根据 Linux 操作系统,图像噪声会随着切片厚度的增加而增加。这些发现可以为 CT 中心的质量控制方法提供有价值的指导,强调需要确定适当的切片厚度以确保准确诊断。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/8cda/10478658/3f49afb69408/JMedLife-16-862-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验