圆锥角膜中常用角膜成像设备的重复性与一致性比较

Comparing Repeatability and Agreement between Commonly Used Corneal Imaging Devices in Keratoconus.

作者信息

Angelo Lize, Gokul Akilesh, McGhee Charles, Ziaei Mohammed

机构信息

Department of Ophthalmology, University of Auckland, Auckland, New Zealand.

出版信息

Optom Vis Sci. 2023 Nov 1;100(11):761-769. doi: 10.1097/OPX.0000000000002079. Epub 2023 Oct 25.

Abstract

SIGNIFICANCE

Keratometry and pachymetry are important for diagnosis and monitoring of keratoconus. Many corneal imaging units exist, but comparison to determine repeatability and agreement is paramount for adequate keratoconus management in primary and secondary care.

PURPOSE

This study aimed to evaluate the repeatability and agreement of a Placido disc-based videokeratoscope (Melbourne, Victoria, Australia) used in primary care with spectral-domain optical coherence tomography (Optopol Technology, Zawiercie, Poland) and Scheimpflug corneal tomography (Oculus, Wetzlar, Germany) in secondary care.

METHODS

This was a prospective single-center study where one eye was randomized to have central corneal thickness;thinnest corneal thickness (TCT); and maximum, mean, steep, and flat keratometry (Kmax, Kmean, Ksteep, and Kflat, respectively), measured with all three devices. Three measurements were completed per device to assess intraobserver repeatability.

RESULTS

A total of 110 eyes from 110 patients with keratoconus were analyzed. Repeatability was best with the Pentacam for central corneal thickness, Kmax, Kmean, Ksteep, and Kflat parameters (precision, 9.21, 0.8, 0.38, 0.52, and 0.58). The Medmont had better repeatability than the Revo with Kmax, Kmean, Ksteep, and Kflat (precision, 1.41,1.35,1.43, and 1.59). Revo had the best repeatability with TCT (precision, 3.81). The intraclass correlation coefficient was >0.94 for all parameters in all devices. Agreement was generally poor between devices. However, there was good agreement between Pentacam and Medmont Kflat measurements ( P > .05).

CONCLUSIONS

Repeatability of keratometry parameters with the Pentacam and Medmont was greater than the Revo, suggesting a lower threshold for detecting change in anterior corneal parameters. The Revo had the greatest repeatability for TCT, suggesting a lower threshold for detecting thinning in disease progression and corneal-crosslinking safety. There was poor agreement between devices, and it is not recommended for them to be used interchangably.

摘要

意义

角膜曲率测量和角膜厚度测量对于圆锥角膜的诊断和监测至关重要。存在多种角膜成像设备,但为了在初级和二级医疗保健中对圆锥角膜进行充分管理,进行比较以确定重复性和一致性至关重要。

目的

本研究旨在评估在初级医疗保健中使用的基于普拉西多盘的视频角膜曲率计(澳大利亚维多利亚州墨尔本)与在二级医疗保健中使用的光谱域光学相干断层扫描(波兰扎维尔采的Optopol Technology)和Scheimpflug角膜断层扫描(德国韦茨拉尔的Oculus)的重复性和一致性。

方法

这是一项前瞻性单中心研究,一只眼睛被随机分配接受所有三种设备测量中央角膜厚度、最薄角膜厚度(TCT)以及最大、平均、陡峭和扁平角膜曲率(分别为Kmax、Kmean、Ksteep和Kflat)。每种设备完成三次测量以评估观察者内重复性。

结果

共分析了110例圆锥角膜患者的110只眼睛。对于中央角膜厚度、Kmax、Kmean、Ksteep和Kflat参数,Pentacam的重复性最佳(精度分别为9.21、0.8、0.38、0.52和0.58)。对于Kmax、Kmean Ksteep和Kflat,Medmont的重复性优于Revo(精度分别为1.41、1.35、1.43和1.59)。Revo在TCT方面具有最佳重复性(精度为3.81)。所有设备中所有参数的组内相关系数均>0.94。各设备之间的一致性总体较差。然而,Pentacam和Medmont的Kflat测量之间具有良好的一致性(P>.05)。

结论

Pentacam和Medmont的角膜曲率测量参数的重复性高于Revo,这表明检测眼前节参数变化的阈值较低。Revo在TCT方面具有最高的重复性,这表明在疾病进展和角膜交联安全性方面检测角膜变薄的阈值较低。各设备之间的一致性较差,不建议它们互换使用。