Dinamap arm and thigh arterial pressure measurement.

作者信息

Goldthorp S L, Cameron A, Asbury A J

出版信息

Anaesthesia. 1986 Oct;41(10):1032-5. doi: 10.1111/j.1365-2044.1986.tb12749.x.

Abstract

Two automatic arterial pressure monitors (Model 1846 P) were used in 30 patients to measure arterial pressures and heart rates in the thigh and arm. Significant differences were found in thigh-arm systolic (p less than 0.001) and diastolic pressures (p less than 0.05). Arm pressures could be predicted from leg pressures and correction factors are suggested. Mean pressures in the thigh approximated more closely to the calculated mean pressures than in the arm. These differences were of little clinical importance.

摘要

使用两台自动动脉压监测仪(型号1846 P)对30例患者测量大腿和手臂的动脉压及心率。发现大腿与手臂的收缩压(p<0.001)和舒张压(p<0.05)存在显著差异。手臂血压可根据腿部血压预测,并给出了校正因子。大腿的平均血压比手臂更接近计算出的平均血压。这些差异临床意义不大。