Goldthorp S L, Cameron A, Asbury A J
Anaesthesia. 1986 Oct;41(10):1032-5. doi: 10.1111/j.1365-2044.1986.tb12749.x.
Two automatic arterial pressure monitors (Model 1846 P) were used in 30 patients to measure arterial pressures and heart rates in the thigh and arm. Significant differences were found in thigh-arm systolic (p less than 0.001) and diastolic pressures (p less than 0.05). Arm pressures could be predicted from leg pressures and correction factors are suggested. Mean pressures in the thigh approximated more closely to the calculated mean pressures than in the arm. These differences were of little clinical importance.