Centre Henri Becquerel, Rouen, France.
Technol Cancer Res Treat. 2023 Jan-Dec;22:15330338231210786. doi: 10.1177/15330338231210786.
The objective was to investigate the possibility of using ExacTrac X-ray (ETX) for 6D image guidance in stereotactic body radiation therapy (SBRT) of bone metastasis and to propose a patient management protocol. The analyses were first obtained from measurements on a pelvic phantom and on 19 patients treated for bone metastasis. The phantom study consisted of applying known offsets and evaluating the ETX level of accuracy, where results were compared with kV-cone beam computed tomography (kV-CBCT). Two groups of patients, 10 spinal and 9 nonspinal SBRT cases, were analyzed to evaluate ETX imaging for different bone localisations. A comparison was made between kV-CBCT and ETX prior to the treatment fractions. During treatments, two other kV-CBCT/ETX image pairs were also acquired and a total of 224 shifts were compared. A second study, using the ETX monitoring module, analyzed the intrafraction motion of 8 other patients. In the phantom study, the root mean square (RMS) of the translational and rotational discrepancies between ETX and kV-CBCT were < 0.6 mm and < 0.4°, respectively. For both groups of patients, the RMS of the discrepancies observed between the two imaging systems were greater than the phantom experiment while still remaining < 1 mm and < 0.7°. In the nonspinal group, three patients (2 scapulas and 1 humerus) did not have consistent shift values with ETX due to a lack of anatomical information. When ETX monitoring was used during irradiation, the setup errors measured were on average less than 1 mm/1°. The results obtained validated the use of ETX for 6D image guidance during bone SBRT. Real-time tracking of the target position improves the accuracy of the irradiation. This strategy allowed for faster correction of out-of-tolerance positioning errors. The registration of bone lesions with poor anatomical information is a limitation of this 2D-kV imaging system.
目的是研究 ExacTrac X 射线(ETX)在立体定向体部放射治疗(SBRT)骨转移中的 6D 图像引导的可能性,并提出患者管理方案。分析首先来自骨盆模型和 19 例骨转移患者的治疗测量。该模型研究包括应用已知偏移量,并评估 ETX 的准确性水平,结果与千伏锥形束计算机断层扫描(kV-CBCT)进行比较。分析了 10 例脊柱和 9 例非脊柱 SBRT 患者,以评估不同骨定位的 ETX 成像。在治疗前对 kV-CBCT 和 ETX 进行了比较。在治疗过程中,还采集了另外两个 kV-CBCT/ETX 图像对,并比较了总共 224 个偏移量。第二项研究使用 ETX 监测模块,分析了另外 8 例患者的分次内运动。在模型研究中,ETX 和 kV-CBCT 之间的平移和旋转差异的均方根(RMS)分别小于 0.6mm 和小于 0.4°。对于两组患者,两个成像系统之间观察到的差异 RMS 值大于模型实验,但仍小于 1mm 和小于 0.7°。在非脊柱组中,由于缺乏解剖学信息,3 例患者(2 例肩胛骨和 1 例肱骨)的 ETX 无一致的偏移值。当在照射过程中使用 ETX 监测时,测量的设置误差平均小于 1mm/1°。结果验证了在骨 SBRT 中使用 ETX 进行 6D 图像引导的可能性。实时跟踪目标位置可提高照射的准确性。这种策略允许更快地纠正超出公差的定位误差。对于解剖学信息较差的骨病变的注册是这个二维千伏成像系统的一个限制。