Comparison of the optical behaviour of five different multifocal diffractive intraocular lenses in a model eye.

机构信息

R & D Department, VSY Biotechnology GMBH, Esslinger Str. 7, 70771, Leinfelden Echterdingen, Germany.

Group of Optics and Visual Perception, Department of Optics, Pharmacology and Anatomy, University of Alicante, Crta San Vicente del Raspeig S/N, 03690, San Vicente del Raspeig, Alicante, Spain.

出版信息

Sci Rep. 2023 Nov 10;13(1):19646. doi: 10.1038/s41598-023-47102-y.

Abstract

The purpose of this study was to investigate and compare the optical performance of five trifocal intraocular lenses (IOLs) following the ISO 11979-2 standards, analysing the impact of tilt and decentration. Five different diffractive trifocal IOLs were evaluated in this experimental study: Acriva Trinova (VSY-Biotechnology) (AT), FineVision HP (PhysIOL) (FVHP), AT LISA tri 839 MP (Zeiss) (ATLT), PanOptix TFNT00 IOL (Alcon) (PO), and Tecnis Synergy (J&J Vision) (TS). In-vitro optical quality analysis of them was performed with the Lambda PMTF system that has an aberration neutral cornea model (Lambda-X Ophthalmics). Measurements were performed on-axis, with 5º of IOL tilt and with 0.5 mm of IOL decentration using 543-nm monochromatic light. Finally, IOL dimensions and diffractive disk profile inspection was performed using the VisIOLA system (Rotlex). On-axis measurements showed a far through-focus MTF > 0.3 at 3 mm aperture, except for TS. FVHP and PO showed better far MTFs for larger apertures (3.75 mm and 4.5 mm) while AT showed good intermediate and near vision for such apertures. With 5º of IOL tilt, the better optical performance at all distances was found with AT for medium-sized pupils (3 mm) and an important reduction of MTF was found for ATLT and PO, especially in the intermediate focus. The induction of 0.5 mm of IOL decentration especially affected the intermediate focus of ATLT and TS and the far focus of FVHP and PO. IOL dimensions and diffractive profile were consistent with those described by the manufacturer. In conclusion, there are differences in the optical performance according to the pupil aperture of the five trifocal IOLs evaluated and this should be considered in clinical practice when selecting the most appropriate implant in each specific case. IOL tilt and decentration can affect significantly in most of the designs evaluated the performance of the IOL at intermediate vision range. It should be noted that measurements were made with an aberration-free cornea, being necessary future studies analysing the impact of different levels of corneal aberrations.

摘要

本研究旨在根据 ISO 11979-2 标准,对 5 种三焦点人工晶状体(IOL)的光学性能进行调查和比较,分析倾斜和偏心的影响。在这项实验研究中评估了 5 种不同的衍射三焦点 IOL:Acriva Trinova(VSY-Biotechnology)(AT)、FineVision HP(PhysIOL)(FVHP)、AT LISA tri 839 MP(蔡司)(ATLT)、PanOptix TFNT00 IOL(爱尔康)(PO)和 Tecnis Synergy(J&J Vision)(TS)。使用具有像差中性角膜模型(Lambda-X Ophthalmics)的 Lambda PMTF 系统对它们进行了体外光学质量分析。在 543nm 单色光下,进行了光轴上测量,IOL 倾斜 5°,IOL 偏心 0.5mm。最后,使用 VisIOLA 系统(Rotlex)进行了 IOL 尺寸和衍射盘轮廓检查。除了 TS 之外,在 3mm 孔径下,所有轴上测量的远焦点调制传递函数(MTF)都大于 0.3。FVHP 和 PO 显示出在较大孔径(3.75mm 和 4.5mm)下更好的远 MTF,而 AT 则显示出在这些孔径下良好的中距离和近视力。在 5°的 IOL 倾斜下,对于中等瞳孔(3mm),AT 在所有距离的光学性能都更好,并且发现 ATLT 和 PO 的 MTF 有重要降低,尤其是在中间焦点。诱导 0.5mm 的 IOL 偏心尤其影响 ATLT 和 TS 的中间焦点以及 FVHP 和 PO 的远焦点。IOL 尺寸和衍射轮廓与制造商描述的一致。总之,根据评估的 5 种三焦点 IOL 的瞳孔孔径,存在光学性能差异,在临床实践中选择每种特定情况下最适合的植入物时应考虑这一点。IOL 倾斜和偏心会显著影响评估的大多数设计的 IOL 中距离视觉范围的性能。需要注意的是,测量是在无像差角膜上进行的,未来的研究需要分析不同水平的角膜像差对其的影响。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/ab19/10638301/1ddf2d97f83e/41598_2023_47102_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验