弹性成像应变率比值在 BI-RADS-US3-4 级良恶性乳腺肿块中的价值。

The value of elastography strain rate ratio in benign and malignant BI-RADS-US 3-4 breast masses.

机构信息

Department of Ultrasonography, The First People's Hospital of Yuhang District, Hangzhou, China.

出版信息

Biomol Biomed. 2024 May 2;24(3):625-632. doi: 10.17305/bb.2023.9878.

Abstract

In the realm of breast diagnostics, Breast Imaging Reporting and Data System (BI-RADS) serves as a systematic framework, guiding a methodical exploration into the nuanced narratives of masses. The current study aims to investigate the value of strain rate ratio (SR) of elastography for benign and malignant breast masses categorized by BI-RADS for Ultrasonography (BI-RADS-US) category 3-4. Ultrasonographic data of 1099 breast masses that underwent both elastography and pathological examination were retrospectively analyzed by non-parametric test, consistency analysis and receiver operating characteristic (ROC) curve sequentially. ROC curve was used to evaluate the diagnostic effect of SR method in different BI-RADS categories of breast lesions. The maximum Youden index obtained from the ROC curve was 0.845. At a cut-off value of 3.57, the SR values' diagnostic sensitivity, specificity, accuracy, positive predictive value and negative predictive value for distinguishing benign and malignant breast lesions were 85.7 %, 98.8 %, 93.6 %, 97.9 % and 91.45 %, respectively. Consistency analysis showed that the consistency between SR value and pathological diagnosis was 93.6 % (κ = 0.864). In addition, the SR values between benign and malignant lesions of BI-RADS 3 and 4 were statistically different (P < 0.001). ROC analysis indicated that the diagnostic area under curve (AUC) for SR value in BI-RADS 3, 4a, 4b and 4c lesions were 0.985, 0.866, 0.793 and 0.916, respectively. In addition, the study observed 58 cases of missed diagnosis and 20 cases of misdiagnosis in evaluating benign and malignant breast lesions. The elastic SR ratio method has a good diagnostic value for the evaluation of breast masses, particularly for lesions with a score of 3-4. The elastic SR ratio method is instrumental in enhancing the accuracy of diagnosis.

摘要

在乳腺诊断领域,乳腺影像报告和数据系统(BI-RADS)为一种系统框架,指导对肿块的细微描述进行有条理的探索。本研究旨在探讨超声 BI-RADS 分类 3-4 类良恶性乳腺肿块的应变率比值(SR)的价值。对 1099 例经超声弹性成像和病理检查的乳腺肿块的超声数据进行回顾性分析,依次采用非参数检验、一致性分析和受试者工作特征(ROC)曲线。ROC 曲线用于评估 SR 法在不同 BI-RADS 类别乳腺病变中的诊断效果。ROC 曲线获得的最大约登指数为 0.845。在 3.57 的截断值下,SR 值鉴别良恶性乳腺病变的诊断灵敏度、特异度、准确度、阳性预测值和阴性预测值分别为 85.7%、98.8%、93.6%、97.9%和 91.45%。一致性分析表明,SR 值与病理诊断的一致性为 93.6%(κ=0.864)。此外,BI-RADS 3 级和 4 级良恶性病变的 SR 值差异有统计学意义(P<0.001)。ROC 分析表明,SR 值在 BI-RADS 3、4a、4b 和 4c 病变中的诊断曲线下面积(AUC)分别为 0.985、0.866、0.793 和 0.916。此外,在评估良恶性乳腺病变时,本研究观察到漏诊 58 例,误诊 20 例。弹性 SR 比值法对乳腺肿块的评估具有良好的诊断价值,尤其对评分 3-4 类的病变具有诊断价值。弹性 SR 比值法有助于提高诊断的准确性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/9bba/11088897/fccd95d3d31b/bb-2023-9878f1.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验