Suppr超能文献

用于持续评估新生儿颅内压的新型 Fontan 测压仪。

New fontanometer for continuous estimation of intracranial pressure in the newborn.

作者信息

Rochefort M J, Rolfe P, Wilkinson A R

出版信息

Arch Dis Child. 1987 Feb;62(2):152-5. doi: 10.1136/adc.62.2.152.

Abstract

Intracranial pressure was estimated by a new pneumatic applanation fontanometer in babies in intensive care. A close correlation with cerebrospinal fluid pressure was found on 35 separate occasions. In 12 control babies the mean (SD) fontanelle pressure was 5.2 (2.2) cm H2O, in eight with a hypoxic-ischaemic encephalopathy it was 12.6 (4.3) cm H2O, and in nine with post-haemorrhagic hydrocephalus it was 10.8 (4.3) cm H2O. In four babies monitored continuously over seven days the correlation remained good. The fontanometer enables pressure to be monitored accurately, continuously, and non-invasively in neonates at risk of raised intracranial pressure.

摘要

在重症监护病房的婴儿中,使用一种新型气动压平式囟门眼压计来估计颅内压。在35次不同的情况下,发现其与脑脊液压力密切相关。在12名对照婴儿中,囟门平均(标准差)压力为5.2(2.2)厘米水柱;在8名患有缺氧缺血性脑病的婴儿中,为12.6(4.3)厘米水柱;在9名患有出血后脑积水的婴儿中,为10.8(4.3)厘米水柱。在4名连续监测7天的婴儿中,相关性仍然良好。这种囟门眼压计能够在有颅内压升高风险的新生儿中准确、连续且无创地监测压力。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/0c12/1778258/2512cecd4a00/archdisch00701-0051-a.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验