Abstract

We evaluated the diagnostic performance of the STANDARD i-Q COVID-19 Ag Test, which was developed to detect viral antigens, using nasal and oral swabs. Sixty positive and 100 negative samples were analyzed. We determined the distribution of the Ct values according to the day of sample collection after symptom onset, the diagnostic performance of the total samples and subgroups separated by Ct value or time of sample collection, and the Ct value at which maximal accuracy was expected. No differences were observed in Ct values, except for the samples obtained on the day of symptom onset. The diagnostic sensitivity and specificity of the oral swabs were 75.0 and 100.0%, respectively, whereas those of the nasal swabs were 85.0 and 98.0%, respectively. The sensitivity was higher in samples with a high viral load collected earlier than those collected later, although the difference was not significant. False-negative results were confirmed in all samples with a Ct value ≥ 30.0. These results indicate that tests using oral and nasal swabs are helpful for diagnosing acute symptomatic cases with suspected high viral loads. Our tests exhibited relatively low sensitivity but high specificity rates, indicating the need to assess negative antigen test results.

摘要

我们评估了用于检测病毒抗原的STANDARD i-Q COVID-19抗原检测试剂盒使用鼻拭子和咽拭子的诊断性能。分析了60份阳性样本和100份阴性样本。我们根据症状出现后样本采集的日期确定了Ct值的分布、按Ct值或样本采集时间划分的总样本及亚组的诊断性能,以及预期最大准确性时的Ct值。除了症状出现当天采集的样本外,未观察到Ct值有差异。咽拭子的诊断敏感性和特异性分别为75.0%和100.0%,而鼻拭子的诊断敏感性和特异性分别为85.0%和98.0%。病毒载量高的样本比病毒载量低的样本在更早采集时敏感性更高,尽管差异不显著。在所有Ct值≥30.0的样本中均确认出现假阴性结果。这些结果表明,使用鼻拭子和咽拭子检测有助于诊断疑似病毒载量高的急性症状病例。我们的检测显示敏感性相对较低但特异性较高,这表明有必要对阴性抗原检测结果进行评估。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/e31e/10814936/9d0551e6fd8b/diagnostics-14-00231-g001a.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验